DOE OSTI · 1660561
AWCC Simulations
Abstract
Neutron well counters provide a means to measure fissile mass by detecting coincident neutrons, which is a unique signature to fission. For coincidence counting, mass is determined using a calibration curve established with items that have the same material characteristics. Multiplicity analysis solves the point model equations for mass using known detector parameters. In both cases, mass measurements may exhibit a bias due to item characteristics and should be corrected for. Experiments can be done to estimate bias and material effects, but they require having both the well counter and the material well defined. In pursuit of estimating measurement bias and uncertainty in a High Efficiency Neutron Counter (HENC), an MCNP simulation was matched to measurements performed with an Active Well Coincidence Counter (AWCC) to de ne a neutron source term. The neutron source term will be used to predict measurement performance of similar material types in a HENC.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Mendoza, Paul Michael, Geist, William H.. 2020-09-02. AWCC Simulations. https://doi.org/10.2172/1660561
Cite the original work for its findings. Save a collection to share your selection of sources.