Continuing our experimental investigation of the fitting accuracy associated with photon correlation spectroscopy, we collect 150 correlograms of light scattered at 90 deg from a thermostated sample of 91-nm-diameter, polystyrene latex spheres in water. The correlograms are taken with two correlators: one with linearly spaced channels and one with geometrically spaced channels. Decay rates are extracted from the single-exponential correlograms with both nonlinear least-squares fits and second-order cumulant fits. We make several statistical comparisons between the two fitting techniques and verify an earlier result that there is no sample-time dependence in the decay rate errors. We find, however, that the two fitting techniques give decay rates that differ by 1 percent.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Shaumeyer, J. N., Briggs, Matthew E., Gammon, Robert W.. 1993-07-20. Statistical fitting accuracy in photon correlation spectroscopy. https://ntrs.nasa.gov/citations/19930068417