Tritium targets for use in solenoidal spectrometers
Nucleon adding and removing reactions are an ideal probe to study the single-particle foundations of nuclear structure. For experiments far away from the valley of stability, one needs to use radioactive beams in inverse kinematics. To maintain excellent resolution, solenoidal spectrometers like HELIOS at Argonne National Laboratory (ANL), ISS at CERN or SOLARIS at the Facility for Rare Isotope Beams (FRIB) are used. Typically, deuterated plastic foils are the targets of choice. Depending on the energy deposited, these targets might degrade quickly in beam. The potential of using tritium-implanted targets in conjunction with the solenoidal spectrometers was realized for access to even more exotic nuclei with outstanding resolving power. While no tritiated polyethylene is commercially available at the moment, tritium-containing titanium targets have been produced in the past. Unfortunately the loading fraction of tritium for the latter was often very low and not reproducible. The progress in the production of tritium-containing targets is presented, including the characterization of the foils.