Similarity Analysis: Comparison of a Standard 60mL With a 50mL Container for MicroCT
A 50mL container with a one inch outer diameter has been proposed for holding specimen materials scanned at LLNL and Tyndall Air Force Research Laboratory (TAFRL). In order to determine whether or not this new container would significantly affect MicroCT results, a similarity analysis procedure developed at LLNL was used. This procedure compares multiple first and second order statistics obtained from CT reconstructed images generated by the MicroCT analysis process, and prescribes similarity constraints as guidance. For this study, water was scanned on the MicroCT system at LLNL, twice in the standard 60mL container and twice in the proposed 50mL container. Data were analyzed following the Standard Operating Procedure (SOP) for MicroCT analysis. Comparison of analytical results shows that the new container does not satisfy all similarity analysis criteria (certain second order statistics are outside of similarity analysis constraints).