Design and validation of a nanosecond short-wave infrared streak camera using gaseous detonation experiments
A custom short-wave infrared (SWIR) galvanometric streak camera was developed to provide nanosecond- scale, time-resolved optical diagnostics. The system was designed as a compact, mechanically driven alternative to traditional tube-based streak cameras, using a galvanometer mirror and multiple-reflection optical path to achieve high temporal resolution while maintaining electronic synchronization for trigger sequencing. Temporal resolution achieved 139.5 ns/px, allowing direct conversion of image-slope gradients to physical wave velocities. The instrument was validated using hydrogen-oxygen-argon detonations within a modular detonation tube, capturing transient emission structures and post-detonation decay behavior with microsecond precision. Although signal levels in filtered configurations were limited by lens transmission and detector sensitivity, the system successfully resolved intensity rise and decay characteristics across 10-60 % argon mixtures. The observed trends demonstrate the feasibility of compact galvanometer-based streak systems for high-speed SWIR imaging in reactive environments!