Decay Curve Correction Analysis Report
The decay curve analysis that is done on the short-lived radionuclide gas samples is used to differentiate between gaseous radionuclides that have the same characteristic gamma decay energy, 511 kiloelectron-volts (keV). A sample of stack gas is isolated and the total counts in the 511 keV peak are counted repeatedly in 10-second intervals to evaluate the decay rate of the sample over time. Analysis of this decay data required a series of steps. First, a raw data report is generated by the gamma acquisition system, based on an analysis template within the acquisition software. The data report file was then loaded into Microsoft Word, and a macro was used to perform minor formatting (remove colons and insert tabs between data columns) to allow analysis within Excel. The file is then saved as a text file at this point. The text file is then uploaded into Excel and a series of macros are used to add labels, calculate radioactive decay constants, and analyze the gamma decay data using linear regression techniques. The analysis template has been used since 1998 for stack 53000303 (TA-53, building 0003, exhaust stack 03) and 2000 for stack 53000702. The overall process, including the gamma report format and the macros used in Word and Excel for processing the report, had remained unchanged until 2015. In October of 2015, staff made a change to the report template in the gamma acquisition software which resulted in an error in the calculations later performed by the Excel macro. This error was not caught until a more in-depth review of the analysis took place regarding 2020 data. This report covers a much more complete review of the issue that occurred regarding the decay curve analysis, a review of the calculations completed to correct the issue, a review of the updated decay curve analysis process, and recommendations for moving forward.