On the Use of Monochromators for the Calibration of AXAF
Data acquired, during the AXAF calibration at MSFC's X-Ray Calibration Facility (XRCF), using monochromators requires special analysis owing to potentially large spatial variations in the monochromators' beams and to higher-order contributions to the spectral content. A description of the monochromators -- the Double-Crystal Monochromator (DCM) and the HIgh-Resolution Erect-Field Spectrometer (HIREFS\TM) reflection-grating monochromator --- is given, followed by a discussion of the spectral and spatial content as monitored by Beam-Normalization Detector (BND) flow proportional counters (FPCs). Emphasis is given to the methodology employed in determining the spatial content through fits to Beam Uniformity (BU) measurements made nonsimultaneously, either during calibration or during pre-calibration source characterization. We present an interpolation scheme which adequately represents the beam properties over a wide range of source conditions.