Directional dependence of equilibrated TLD-400 chips in multiple radiation fields
Fielding of thermoluminescent dosimeters (TLDs) for measurement of photon radiation dose in experiments is the standard practice for γ-irradiation facilities, pulsed power x-ray facilities, and reactor facilities at Sandia National Laboratories (SNL). Due to the high-dose experimental conditions and the mixed ( 1 n, γ) fields in these facilities, SNL radiation metrologists have historically used CaF 2 :Mn TLDs (also known as TLD-400). Recent inquiries to the radiation metrology staff have raised concerns that the aluminum-equilibrated TLD-400 chips may exhibit a directionally dependent response. The metrologists were asked whether the dose measured by the chip may be impacted by the angle of incidence on the equilibrated TLD. To provide a thorough answer to this query, a set of adjoint Monte Carlo radiation transport calculations was performed for three different equilibrated TLDs as well as bare TLD chips using the Integrated Tiger Series (ITS) code. The key feature of the adjoint calculations performed in this study is that each photon escaping the modeled geometry was tallied into angular bins to provide information on both the energy and angular dependence of the equilibrated chip. After the adjoint Monte Carlo calculations were completed, the resulting energy-dependent response function for each angular bin was convolved with multiple photon energy spectra representing various radiation facilities at SNL. The radiation facilities selected for analysis span a range of photon energies from approximately 1 keV up to approximately 20 MeV. Thus, the presented results are applicable to a wide variety of radiation facilities around the world. Although the bare TLD-400 chip was expected to display the largest variation due to the photon angle of impact on the dosimeter, the dosimeter with the thinnest aluminum equilibrator (SNL thin equilibrated TLD-400) was determined to have the biggest differential between the impact angle with the maximum dose ( D max ) and the impact angle with the minimum dose ( D min ). However, the SNL normal equilibrated TLD-400 chip demonstrated a dramatic reduction in that differential between maximum and minimum dose angles. The reduction in this differential is one of the dominant factors in experimenters’ choice to field these dosimeters at SNL radiation facilities. The results from the PNNL (Hanford) energy-flattening field capsule are consistent with the previous publications. The differential dose responses due to the photon impact angle indicate that experimenters should strive to field their TLD-400 dosimeters in a consistent manner to avoid additional uncertainty in the measurements based on dosimeter orientation.