Spectroscopic detection of biological NO with a quantum cascade laser

Two configurations of a continuous wave quantum cascade distributed feedback laser-based gas sensor for the detection of NO at a parts per billion (ppb) concentration level, typical of biomedical applications, have been investigated. The laser was operated at liquid nitrogen temperature near lambda = 5.2 microns. In the first configuration, a 100 m optical path length multi-pass cell was employed to enhance the NO absorption. In the second configuration, a technique based on cavity-enhanced spectroscopy (CES) was utilized, with an effective path length of 670 m. Both sensors enabled simultaneous analysis of NO and CO2 concentrations in exhaled air. The minimum detectable NO concentration was found to be 3 ppb with a multi-pass cell and 16 ppb when using CES. The two techniques are compared, and potential future developments are discussed.

NASA Discipline Environmental Health

Pulmonary function in microgravity: Spacelab 4 and beyond

This paper refers principally to the composition gradient of gases within the lung in various conditions of gravity, as revealed by exhaled breath. A rapid gas analyzer-based system has been developed for tests in Spacelab 4. The test sequence and expected results are presented.

Flight Experiment