System Response Testing of HEAF Kevex 130kV Microfocus Source and Thales Flashscan Panel I. Step-wedge Measurements
In the Small Sample DR Testing series, a few questions have arisen when comparing modeled results to experimental data. It is necessary to evaluate these questions in order to validate the modeling assumptions and to ensure the validity of the test procedure in the Small Sample DR Testing. Open Questions: (1) The decay time of the scintillator elements is unknown. (2) The thickness of the Beryllium window on the source is unknown. (3) The residual effects of beam angle and scattering within the detector assembly are unknown. (4) The spectral response of the scintillator and digitizing panel are not known. (5) The spectral content of the source is unknown. Item 1 will be addressed in separate tests by measuring the output of the panel at various intervals after the source has stopped illuminating the panel. Item 2 will be determined by finding the manufacturer specifications. Item 3 will be addressed in separate tests. We cannot easily address items 4 and 5 independently. Measuring the spectral responsivity of the detector directly would require the use of a calibrated reference source, which we lack. Similarly, measuring the spectral content of the source requires spectrally and spatially calibrated detectors, which we also lack. However, we can address the combination of items 1 and 2 to deduce the product of source intensity times detector response. By measuring the X-ray transmittance of samples of known thickness, material composition and density (step wedges), we can infer the desired product and so validate and improve computer models of our system.