Lanthanum Bromide Detector System for Mu2e Stopping Target Monitor
The stopping target monitor (STM) in the Mu2e experiment is used to normalize the muon decay branching ratio in the Mu2e experiment, by estimating the number of muons that have been captured on the Mu2e nuclear target. The STM measures the distinctive muonic atomic transition lines and muonic nuclear excitations produced in the Aluminum Stopping Target of Mu2e. These signals must be disentangled in the presence of high rate Bremsstrahlung and other backgrounds produced in the target. An HPGe detector with fine energy resolution, ~0.1% will be employed but has limited count rate capability, ~20kcps. Time-to-Decision, that is the time to make an estimate of the experiments observed luminosity, is inversely proportional to the signal rate squared. For this reason the STM will also use a Lanthanum Bromide(LaBr3) detector, having ~2% energy resolution but capable of ~1 Mcps counting rate. This higher rate will allow not only an orthogonal verification of the HPGe luminosity measurements, but will also yield much finer luminosity measurements as a function of time because of its significantly higher counting rate capabilities. Geant4 simulations are presented comparing the performance of different sized, LaBr3 crystals. The results highly favor the use of the commercially available right circular cylindrical, 3x3-in. detector. Modifications to the STM design are discussed in order to take full advantage of the LaBr3 high rate and fast decision making capabilities. A preliminary design will be presented for a custom read-out system needed to fulfill these requirements.