Fading study of Harshaw LiF:Mg,Ti thermoluminescence dosimeters exposed to neutron doses
Sensitivity and signal fading were tested in Harshaw LiF:Mg,Ti thermoluminescence dosimeters (TLDs) exposed to neutron and beta/photons doses. A significant difference in the signal fading rate was observed for TLDs exposed to neutrons compared with TLDs exposed to betas. This difference may have been related to differences in glow curves for dosimeters exposed to different beams and to differences in glow curve evolution after exposure. A fading-correction algorithm was developed to correct the cumulative sensitivity + signal fading of TL doses for the fading time t using the formulae: y = 0.147∗exp(-t/4.36) + 0.308∗exp(-t/34.1) + 0.73 for neutrons, and: y = 0.132∗exp(-t/10.72) + 0.174∗exp(-t/31.5) + 0.8 for betas/photons. The formula was validated using the dose recovery test, where the average deviation of fading-corrected doses from delivery doses was within 1%. In conclusion, the proposed fading-correction approaches may significantly improve the accuracy of Harshaw TL dosimetry with LiF:Mg,Ti cards for both neutron and beta/gamma irradiation.