Volatile Organic Analyzer (VOA) in 2006: Repair, Revalidation, and Restart of Elektron Even
The Volatile Organic Analyzer (VOA) had been providing valuable data on trace contaminants in the atmosphere of the International Space Station (ISS) from January 2002 through May 2003. Component temperature errors, detected by the VOA s software, shut down the unit in May 2003, but in early 2005 on orbit diagnostics verified fuse failures had disabled both VOA channels. An in-flight maintenance (IFM) session in December 2005 returned the VOA to an operational mode by January 2006. This paper will present the on-orbit data from 2006 that were used to revalidate the VOA, and provide an overview of the VOA s contributions during the Elecktron contingency event that occurred on ISS in September 2006.