Analysis of the new polarimeter for the Marshall Space Flight Center vector magnetograph
The magnetograph was upgraded in both electronic control of the magnetograph hardware and in the polarization optics. The problems associated with the orignal polarimeter were: (1) field of view errors associated with the natural birefringence of the KD*P crystals; (2.) KD*P electrode failure due to the halfwave dc voltage required in one of the operational sequences; and (3) breakdown of the retardation properties of some KD*Ps when exposed to a zero to halfwave modulation (DC) scheme. The new polarimeter gives up the flexibility provided by two variable waveplates to adjust the retardances of the optics for a particular polarization measurement, but solves the problems associated with the original polarimeter. With the addition of the quartz quarterwave plates, a new optical alignment was developed to allow the remaining KD*P to correct for errors in the waveplates. The new optical alignment of the polarimeter is prescribed. The various sources of error, and how those errors are minimized so that the magnetograph can look at the transverse field in real time are discussed.