Engineering Papers⌕ Search

SEARCH · Engineering Papers

Results for “Linear accelerator”

Search indexed NASA NTRS and DOE OSTI research on propulsion, heat transfer, battery materials and energy systems. Follow report and document links to the original sources.

Quote a phrase for an exact phrase match. Source license links do not imply unrestricted reuse.

At least 991 records · Page 55

TH-AB-BRB-01: Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation

Current state-of-the art digital C-arm medical linear accelerators are capable of delivering radiation treatments with high level of automation, which affords coordinated motions of gantry, couch, and multileaf collimator (MLC) with dose rate modulations. The new machine capacity has shown the potential to bring substantially improved radiation dosimetry and/or delivery efficiency to many challenging diseases. Combining an integrated beam orientation optimization algorithm with automated machine navigation, markedly improved dose conformity has been achieved using 4ρ therapy. Trajectory modulated radiation therapy (TMAT) can be used to deliver highly conformal dose to partial breast or to carve complex dose distribution for therapy involving extended volumes such as total marrow and total lymph node treatment. Dynamic electron arc radiotherapy (DEAR) not only overcomes the deficiencies of conventional electron therapy in dose conformity and homogeneity but also achieves so without patient-specific shields. The combination of MLC and couch tracking provides improved motion management of thoracic and abdominal tumors. A substantial body of work has been done in these technological advances for clinical translation. The proposed symposium will provide a timely review of these exciting opportunities. Learning Objectives: Recognize the potential of using digitally controlled linacs for clinically significant improvements in delivered dose distributions for various treatment sites. Identify existing approaches to treatment planning, optimization and delivery for treatment techniques utilizing the advanced functions of digital linacs and venues for further development and improvement. Understand methods for testing and validating delivery system performance. Identify tools available on current delivery systems for implementation and control for such treatments. Obtain the update in clinical applications, trials and regulatory approval. K. Sheng, NIH U19AI067769, NIH R43CA183390, NIH R01CA188300, Varian Medical Systems V. Yu, Varian Medical Systems, AAPM Summer Undergraduate Fellowship, NSF graduate fellowship S. Nill, Elekta AB. Cancer Research UK under Programme C33589/A19727, NIHR Biomedical Research Centre at The Royal Marsden and The Institute of Cancer Research.

60 APPLIED LIFE SCIENCES↗

TH-AB-BRB-02: ET-Reloaded

Current state-of-the art digital C-arm medical linear accelerators are capable of delivering radiation treatments with high level of automation, which affords coordinated motions of gantry, couch, and multileaf collimator (MLC) with dose rate modulations. The new machine capacity has shown the potential to bring substantially improved radiation dosimetry and/or delivery efficiency to many challenging diseases. Combining an integrated beam orientation optimization algorithm with automated machine navigation, markedly improved dose conformity has been achieved using 4ρ therapy. Trajectory modulated radiation therapy (TMAT) can be used to deliver highly conformal dose to partial breast or to carve complex dose distribution for therapy involving extended volumes such as total marrow and total lymph node treatment. Dynamic electron arc radiotherapy (DEAR) not only overcomes the deficiencies of conventional electron therapy in dose conformity and homogeneity but also achieves so without patient-specific shields. The combination of MLC and couch tracking provides improved motion management of thoracic and abdominal tumors. A substantial body of work has been done in these technological advances for clinical translation. The proposed symposium will provide a timely review of these exciting opportunities. Learning Objectives: Recognize the potential of using digitally controlled linacs for clinically significant improvements in delivered dose distributions for various treatment sites. Identify existing approaches to treatment planning, optimization and delivery for treatment techniques utilizing the advanced functions of digital linacs and venues for further development and improvement. Understand methods for testing and validating delivery system performance. Identify tools available on current delivery systems for implementation and control for such treatments. Obtain the update in clinical applications, trials and regulatory approval. K. Sheng, NIH U19AI067769, NIH R43CA183390, NIH R01CA188300, Varian Medical Systems V. Yu, Varian Medical Systems, AAPM Summer Undergraduate Fellowship, NSF graduate fellowship S. Nill, Elekta AB. Cancer Research UK under Programme C33589/A19727, NIHR Biomedical Research Centre at The Royal Marsden and The Institute of Cancer Research.

60 APPLIED LIFE SCIENCES↗

TH-AB-BRB-03: 4n Radiotherapy

Current state-of-the art digital C-arm medical linear accelerators are capable of delivering radiation treatments with high level of automation, which affords coordinated motions of gantry, couch, and multileaf collimator (MLC) with dose rate modulations. The new machine capacity has shown the potential to bring substantially improved radiation dosimetry and/or delivery efficiency to many challenging diseases. Combining an integrated beam orientation optimization algorithm with automated machine navigation, markedly improved dose conformity has been achieved using 4ρ therapy. Trajectory modulated radiation therapy (TMAT) can be used to deliver highly conformal dose to partial breast or to carve complex dose distribution for therapy involving extended volumes such as total marrow and total lymph node treatment. Dynamic electron arc radiotherapy (DEAR) not only overcomes the deficiencies of conventional electron therapy in dose conformity and homogeneity but also achieves so without patient-specific shields. The combination of MLC and couch tracking provides improved motion management of thoracic and abdominal tumors. A substantial body of work has been done in these technological advances for clinical translation. The proposed symposium will provide a timely review of these exciting opportunities. Learning Objectives: Recognize the potential of using digitally controlled linacs for clinically significant improvements in delivered dose distributions for various treatment sites. Identify existing approaches to treatment planning, optimization and delivery for treatment techniques utilizing the advanced functions of digital linacs and venues for further development and improvement. Understand methods for testing and validating delivery system performance. Identify tools available on current delivery systems for implementation and control for such treatments. Obtain the update in clinical applications, trials and regulatory approval. K. Sheng, NIH U19AI067769, NIH R43CA183390, NIH R01CA188300, Varian Medical Systems V. Yu, Varian Medical Systems, AAPM Summer Undergraduate Fellowship, NSF graduate fellowship S. Nill, Elekta AB. Cancer Research UK under Programme C33589/A19727, NIHR Biomedical Research Centre at The Royal Marsden and The Institute of Cancer Research.

60 APPLIED LIFE SCIENCES↗

TH-AB-BRB-05: Using a Research Real-Time Control Interface to Go Beyond Dynamic MLC Tracking

Current state-of-the art digital C-arm medical linear accelerators are capable of delivering radiation treatments with high level of automation, which affords coordinated motions of gantry, couch, and multileaf collimator (MLC) with dose rate modulations. The new machine capacity has shown the potential to bring substantially improved radiation dosimetry and/or delivery efficiency to many challenging diseases. Combining an integrated beam orientation optimization algorithm with automated machine navigation, markedly improved dose conformity has been achieved using 4ρ therapy. Trajectory modulated radiation therapy (TMAT) can be used to deliver highly conformal dose to partial breast or to carve complex dose distribution for therapy involving extended volumes such as total marrow and total lymph node treatment. Dynamic electron arc radiotherapy (DEAR) not only overcomes the deficiencies of conventional electron therapy in dose conformity and homogeneity but also achieves so without patient-specific shields. The combination of MLC and couch tracking provides improved motion management of thoracic and abdominal tumors. A substantial body of work has been done in these technological advances for clinical translation. The proposed symposium will provide a timely review of these exciting opportunities. Learning Objectives: Recognize the potential of using digitally controlled linacs for clinically significant improvements in delivered dose distributions for various treatment sites. Identify existing approaches to treatment planning, optimization and delivery for treatment techniques utilizing the advanced functions of digital linacs and venues for further development and improvement. Understand methods for testing and validating delivery system performance. Identify tools available on current delivery systems for implementation and control for such treatments. Obtain the update in clinical applications, trials and regulatory approval. K. Sheng, NIH U19AI067769, NIH R43CA183390, NIH R01CA188300, Varian Medical Systems V. Yu, Varian Medical Systems, AAPM Summer Undergraduate Fellowship, NSF graduate fellowship S. Nill, Elekta AB. Cancer Research UK under Programme C33589/A19727, NIHR Biomedical Research Centre at The Royal Marsden and The Institute of Cancer Research.

60 APPLIED LIFE SCIENCES↗

SU-E-T-781: Using An Electronic Portal Imaging Device (EPID) for Correlating Linac Photon Beam Energies

Purpose: Electronic portal imaging devices (EPIDs) have proven to be useful for measuring several parameters of interest in linear accelerator (linac) quality assurance (QA). The purpose of this project was to evaluate the feasibility of using EPIDs for determining linac photon beam energies. Methods: Two non-clinical Varian TrueBeam linacs (Varian Medical Systems, Palo Alto, CA) with 6MV and 10MV photon beams were used to perform the measurements. The linacs were equipped with an amorphous silicon based EPIDs (aSi1000) that were used for the measurements. We compared the use of flatness versus percent depth dose (PDD) for predicting changes in linac photon beam energy. PDD was measured in 1D water tank (Sun Nuclear Corporation, Melbourne FL) and the profiles were measured using 2D ion-chamber array (IC-Profiler, Sun Nuclear) and the EPID. Energy changes were accomplished by varying the bending magnet current (BMC). The evaluated energies conformed with the AAPM TG142 tolerance of ±1% change in PDD. Results: BMC changes correlating with a ±1% change in PDD corresponded with a change in flatness of ∼1% to 2% from baseline values on the EPID. IC Profiler flatness values had the same correlation. We observed a similar trend for the 10MV beam energy changes. Our measurements indicated a strong correlation between changes in linac photon beam energy and changes in flatness. For all machines and energies, beam energy changes produced change in the uniformity (AAPM TG-142), varying from ∼1% to 2.5%. Conclusions: EPID image analysis of beam profiles can be used to determine linac photon beam energy changes. Flatness-based metrics or uniformity as defined by AAPM TG-142 were found to be more sensitive to linac photon beam energy changes than PDD. Research funding provided by Varian Medical Systems. Dr. Sasa Mutic receives compensation for providing patient safety training services from Varian Medical Systems, the sponsor of this study.

61 RADIATION PROTECTION AND DOSIMETRY↗

SU-F-T-526: A Comparative Study On Gating Efficiency of Varian RPM Device and Calypso System

Purpose: In general, the linear accelerator is gated using respiratory signal obtained by way of external sensors to account for the breathing motion during radiotherapy. One of the commonly used gating devices is the Varian RPM device. Calypso system that uses electromagnetic tracking of implanted or surface transponders could also be used for gating. The aim of this study is to compare the gating efficiency of RPM device and the calypso system by phantom studies. Methods: An ArcCheck insert was used as the phantom with a Gafchromic film placed in its holder. The ArcCheck insert was placed on a Motion Sim platform and moved in the longitudinal direction simulating a respiratory motion with a period of 5 seconds and amplitude of ±6mm. The Gafchromic film was exposed to a 2 × 2cm{sup 2} field, i) with the phantom static, ii) phantom moving but ungated iii) gated with gating window of 2mm and 3mm. This was repeated with Calypso system using surface transponders with the same gating window. The Gafchromic films were read with an EPSON 11000 flatbed scanner and analysed with ‘Medphysto’ software. Results: The full width at half maximum (FWHM) as measured with film at the level of the film holder was 1.65cm when the phantom was static. FWHM measured with phantom moving and without gating was 1.16 cm and penumbra was 7 mm (80–20%) on both sides. When the beam was gated with 2 mm gating window the FWHM was 1.8 cm with RPM device and 1.9 cm with Calypso. Similarly, when the beam was gated with 3 mm window, the FWHM was 1.9cm with RPM device and 2cm with Calypso. Conclusion: This work suggests that the gating efficiency of RPM device is better than that of the Calypso with surface transponder, with reference to the latency in gating.

60 APPLIED LIFE SCIENCES↗

TH-AB-BRB-04: Quality Assurance for Advanced Digital Linac Implementations

Current state-of-the art digital C-arm medical linear accelerators are capable of delivering radiation treatments with high level of automation, which affords coordinated motions of gantry, couch, and multileaf collimator (MLC) with dose rate modulations. The new machine capacity has shown the potential to bring substantially improved radiation dosimetry and/or delivery efficiency to many challenging diseases. Combining an integrated beam orientation optimization algorithm with automated machine navigation, markedly improved dose conformity has been achieved using 4ρ therapy. Trajectory modulated radiation therapy (TMAT) can be used to deliver highly conformal dose to partial breast or to carve complex dose distribution for therapy involving extended volumes such as total marrow and total lymph node treatment. Dynamic electron arc radiotherapy (DEAR) not only overcomes the deficiencies of conventional electron therapy in dose conformity and homogeneity but also achieves so without patient-specific shields. The combination of MLC and couch tracking provides improved motion management of thoracic and abdominal tumors. A substantial body of work has been done in these technological advances for clinical translation. The proposed symposium will provide a timely review of these exciting opportunities. Learning Objectives: Recognize the potential of using digitally controlled linacs for clinically significant improvements in delivered dose distributions for various treatment sites. Identify existing approaches to treatment planning, optimization and delivery for treatment techniques utilizing the advanced functions of digital linacs and venues for further development and improvement. Understand methods for testing and validating delivery system performance. Identify tools available on current delivery systems for implementation and control for such treatments. Obtain the update in clinical applications, trials and regulatory approval. K. Sheng, NIH U19AI067769, NIH R43CA183390, NIH R01CA188300, Varian Medical Systems V. Yu, Varian Medical Systems, AAPM Summer Undergraduate Fellowship, NSF graduate fellowship S. Nill, Elekta AB. Cancer Research UK under Programme C33589/A19727, NIHR Biomedical Research Centre at The Royal Marsden and The Institute of Cancer Research.

60 APPLIED LIFE SCIENCES↗

SU-E-T-533: Evaluation of Dose Calculation Accuracy for Small Elongated Targets On the Edge Linac

Purpose: To evaluate output factors and dose calculation accuracy on a novel SRS linear accelerator, the Edge (Varian), for treatments of small, elongated targets using flattening filter free (FFF) beam. Methods: Total scatter/output factors (OF’s) for 24 elongated, small, high definition multi-leaf collimator (HDMLC)-defined fields were measured on the Edge machine using 6X FFF beam. 3 detectors were used in water tank: CC01 ion chamber (active volume 10cc), stereotactic photon diode (SFD) (active diameter 0.6mm, active thickness 0.06mm), Edge detector (active volume 0.0019cc). The 24 MLC apertures have widths ranging from 5 to 20mm and length/width ratio from 0.25 to 5. Readings were cross calibrated with CC04 at field size 3×3 cm. A beam model was developed using commissioning measurements for treatment planning in Eclipse (AAA, version 11). One representative patient case (IMRT, target volume 0.2cc, 4×4×14mm) was calculated using AAA 11 and delivered on the Edge. Results: Due to volume averaging effects, CC01 readings were 11.2±0.9% lower than SFD readings for 5mm field sizes. The Edge diode showed a uniform over-response of 2.6±0.7% compared to SFD. Calculation using AAA v11 showed the best agreement with SFD measurements (2.4±1.7% lower than SFD). The largest difference between AAA v11 and SFD occurs at 5mm field sizes. For the patient plan, dose delivered on Edge was measured to be 2.2% higher than AAA v11 calculation. Conclusion: Cross-calibrated SFD output measurements presented the best agreement with commissioned AAA v11 beam model. Field sizes smaller than 1cm posed challenges to both the detectors and the calculation algorithm. For the representative patient with small elongated target, AAA v11 and measurements agreed within ~2% on the Edge linac. Although encouraging, a more comprehensive study is required to validate the overall algorithmic accuracy.

60 APPLIED LIFE SCIENCES↗

Radiotherapy beyond cancer: Target localization in real-time MRI and treatment planning for cardiac radiosurgery

Purpose: Atrial fibrillation (AFib) is the most common cardiac arrhythmia that affects millions of patients world-wide. AFib is usually treated with minimally invasive, time consuming catheter ablation techniques. While recently noninvasive radiosurgery to the pulmonary vein antrum (PVA) in the left atrium has been proposed for AFib treatment, precise target location during treatment is challenging due to complex respiratory and cardiac motion. A MRI linear accelerator (MRI-Linac) could solve the problems of motion tracking and compensation using real-time image guidance. In this study, the authors quantified target motion ranges on cardiac magnetic resonance imaging (MRI) and analyzed the dosimetric benefits of margin reduction assuming real-time motion compensation was applied. Methods: For the imaging study, six human subjects underwent real-time cardiac MRI under free breathing. The target motion was analyzed retrospectively using a template matching algorithm. The planning study was conducted on a CT of an AFib patient with a centrally located esophagus undergoing catheter ablation, representing an ideal case for cardiac radiosurgery. The target definition was similar to the ablation lesions at the PVA created during catheter treatment. Safety margins of 0 mm (perfect tracking) to 8 mm (untracked respiratory motion) were added to the target, defining the planning target volume (PTV). For each margin, a 30 Gy single fraction IMRT plan was generated. Additionally, the influence of 1 and 3 T magnetic fields on the treatment beam delivery was simulated using Monte Carlo calculations to determine the dosimetric impact of MRI guidance for two different Linac positions. Results: Real-time cardiac MRI showed mean respiratory target motion of 10.2 mm (superior–inferior), 2.4 mm (anterior–posterior), and 2 mm (left–right). The planning study showed that increasing safety margins to encompass untracked respiratory motion leads to overlapping structures even in the ideal scenario, compromising either normal tissue dose constraints or PTV coverage. The magnetic field caused a slight increase in the PTV dose with the in-line MRI-Linac configuration. Conclusions: The authors’ results indicate that real-time tracking and motion compensation are mandatory for cardiac radiosurgery and MRI-guidance is feasible, opening the possibility of treating cardiac arrhythmia patients completely noninvasively.

60 APPLIED LIFE SCIENCES↗

SU-F-T-288: Impact of Trajectory Log Files for Clarkson-Based Independent Dose Verification of IMRT and VMAT

Purpose: To investigate the effect of the trajectory files from linear accelerator for Clarkson-based independent dose verification in IMRT and VMAT plans. Methods: A CT-based independent dose verification software (Simple MU Analysis: SMU, Triangle Products, Japan) with a Clarksonbased algorithm was modified to calculate dose using the trajectory log files. Eclipse with the three techniques of step and shoot (SS), sliding window (SW) and Rapid Arc (RA) was used as treatment planning system (TPS). In this study, clinically approved IMRT and VMAT plans for prostate and head and neck (HN) at two institutions were retrospectively analyzed to assess the dose deviation between DICOM-RT plan (PL) and trajectory log file (TJ). An additional analysis was performed to evaluate MLC error detection capability of SMU when the trajectory log files was modified by adding systematic errors (0.2, 0.5, 1.0 mm) and random errors (5, 10, 30 mm) to actual MLC position. Results: The dose deviations for prostate and HN in the two sites were 0.0% and 0.0% in SS, 0.1±0.0%, 0.1±0.1% in SW and 0.6±0.5%, 0.7±0.9% in RA, respectively. The MLC error detection capability shows the plans for HN IMRT were the most sensitive and 0.2 mm of systematic error affected 0.7% dose deviation on average. Effect of the MLC random error did not affect dose error. Conclusion: The use of trajectory log files including actual information of MLC location, gantry angle, etc should be more effective for an independent verification. The tolerance level for the secondary check using the trajectory file may be similar to that of the verification using DICOM-RT plan file. From the view of the resolution of MLC positional error detection, the secondary check could detect the MLC position error corresponding to the treatment sites and techniques. This research is partially supported by Japan Agency for Medical Research and Development (AMED)

60 APPLIED LIFE SCIENCES↗

SU-F-T-383: Robustness for Patient Setup Error in Total Body Irradiation Using Volumetric Modulated Arc Therapy (VMAT)

Purpose: Total body irradiation (TBI) and total marrow irradiation (TMI) using Tomotherapy have been reported. A gantry-based linear accelerator uses one isocenter during one rotational irradiation. Thus, 3–5 isocenter points should be used for a whole plan of TBI-VMAT during smoothing out the junctional dose distribution. IGRT provides accurate and precise patient setup for the multiple junctions, however it is evident that some setup errors should occur and affect accuracy of dose distribution in the area. In this study, we evaluated the robustness for patient’s setup error in VMAT-TBI. Methods: VMAT-TBI Planning was performed in an adult whole-body human phantom using Eclipse. Eight full arcs with four isocenter points using 6MV-X were used to cover the entire whole body. Dose distribution was optimized using two structures of patient’s body as PTV and lung. The two arcs were shared with one isocenter and the two arcs were 5 cm-overlapped with the other two arcs. Point absolute dose using ionization-chamber and planer relative dose distribution using film in the junctional regions were performed using water-equivalent slab phantom. In the measurements, several setup errors of (+5∼−5mm) were added. Results: The result of the chamber measurement shows the deviations were within ±3% when the setup errors were within ±3 mm. In the planer evaluation, the pass ratio of gamma evaluation (3%/2mm) shows more than 90% if the errors within ±3 mm. However, there were hot/cold areas in the edge of the junction even with acceptable gamma pass ratio. 5 mm setup error caused larger hot and cold areas and the dosimetric acceptable areas were decreased in the overlapped areas. Conclusion: It can be clinically acceptable for VMAT-TBI when patient setup error is within ±3mm. Averaging effects from patient random error would be helpful to blur the hot/cold area in the junction.

60 APPLIED LIFE SCIENCES↗

SU-F-T-529: Dosimetric Investigation of a Rotating Gamma Ray System for ImagedGuided Modulated Arc Radiotherapy

Purpose: Because of their effectiveness and efficiency, rotational arc radiotherapy (MAT) techniques have been developed on both specialty machines such as Tomotherapy and conventional linear accelerators. This work investigates a new rotating Gamma therapy system for image guided MAT and SBRT of intra/extracranial tumors. Methods: The CyberMAT system (Cyber Medical Corp., China) consists of a ring gantry with a gamma source (effective source size 1cm and 1.7cm respectively), a 120-leaf MLC, a kV CBCT and an EPID. The treatment couch provides 6-degrees-of-freedom motion compensation and the kV CBCT system has a spatial resolution of 0.4mm for target localization. The maximum dose rate is >4.0 Gy/min and the maximum field size is 40cm × 40cm. Monte Carlo simulations were used to compute dose distributions and compare with measurements. A retrospective study of 125 previously treated SBRT patients was performed to evaluate the dosimetric characteristics of CyberMAT in comparison with existing VMAT systems. Results: Monte Carlo results confirmed the CyberMAT design parameters including output factors and 3D dose distributions. Its beam penumbra is 5mm to 10mm for field sizes 1cm to 10cm, respectively and its isocenter accuracy is <0.5mm. Compared to the 6 MV photons of Tomotherapy and conventional linacs, Cobalt beams produce lower-energy secondary electrons that exhibit better dose properties in low-density lung tissues. Cobalt beams are ideal for peripheral lung tumors with half-arc arrangements to spare the opposite lung and other critical structures. Superior dose distributions have been obtained for brain, head and neck, breast, spine and lung tumors with half/full arc arrangements. Conclusion: Because of the unique dosimetric properties of Cobalt sources and its accurate stereotaxy/dose delivery CyberMAT is ideally suited for image guided MAT and SBRT. Full-arc arrangements are superior for brain and H&N treatments while half-arc arrangements produce best dose distributions for thoracic tumors.

60 APPLIED LIFE SCIENCES↗

WE-AB-207A-11: Respiratory Motion Guided 4DCBCT - A Step Towards Controlling 4DCBCT Image Quality

Purpose: We have developed a method, called respiratory motion guided 4DCBCT (RMG-4DCBCT), in which the gantry speed and projection frequency are varied in response to the patient’s real-time respiratory signal to eliminate streaking artifacts and to suppress duplicate projections in 4DCBCT images. In 2015, we realized RMG-4DCBCT on an Elekta Synergy linear accelerator with a mechanical relay to suppress projections and a potentiometer to adjust the gantry speed in response to the patient’s real-time respiratory signal. The aim of this study was to analyse the image quality to determine what can and cannot be controlled. Methods: Using RMG-4DCBCT, we acquired 40 (RMG-4DCBCT-40) and 60 (RMG-4DCBCT-60) equally spaced projections per respiratory phase of the CIRS dynamic thorax phantom with breathing periods from 2s to 8s and two breathing traces from lung cancer patients. The contrast to noise ratio (CNR) and edge response width (ERW) were used to compare image quality between RMG-4DCBCT and conventional 4DCBCT. Results: Regardless of the breathing period, for RMG-4DCBCT, the CNR is approximately 7 and 9 with RMG-4DCBCT-40 and RMG-4DCBCT-60 respectively. Conventional 4DCBCT has a CNR dropping from 20 down to 6 as the breathing period drops from 2s to 8s. With RMG-4DCBCT, the ERW, in the direction of phantom motion, ranges from 2.1mm to 2.5mm as the breathing period drops from 2s to 8s which compares to a higher range of 2.0mm to 2.5mm with conventional 4DCBCT. Images with similar quality to conventional 4DCBCT can be acquired with RMG-4DCBCT-40 which has a 70% reduction in imaging dose. Conclusion: The image contrast can be controlled with RMG-4DCBCT regardless of the patients breathing rate. However, although the image sharpness is better with RMG-4DCBCT, image sharpness has a small dependence on the breathing period; the accuracy of registration and segmentation will therefore vary with the patient’s breathing period. This project was supported by a National Health and Medical Research Council (NHMRC) project grant 1034060 and Cancer Australia grant number 1084566.

60 APPLIED LIFE SCIENCES↗

SU-F-T-524: Investigation of the Dosimertric Benefits of Interchangeable Source Size of a Novel Rotating Gamma System

Purpose: Tremendous technological developments were made for conformal therapy techniques with linear accelerators, while less attention was paid to cobalt-60 units. The aim of the current study is to explore the dosimetric benefits of a novel rotating gamma ray system enhanced with interchangeable source sizes and multi-leaf collimator (MLC). Material and Methods: CybeRT is a novel rotating gamma ray machine with a ring gantry that ensures an iso-center accuracy of less than 0.3 mm. The new machine has a 70cm source axial distance allowing for improved penumbra compared to conventional machines. MCBEAM was used to simulate Cobalt-60 beams from the CybeRT head, while the MCPLAN code was used for modeling the MLC and for phantom/patient dose calculation. The CybeRT collimation will incorporate a system allowing for interchanging source sizes. In this work we have created phase space files for 1cm and 2cm source sizes. Evaluation of the system was done by comparing CybeRT beams with the 6MV beams in a water phantom and in patient geometry. Treatment plans were compared based on isodose distributions and dose volume histograms. Results: Profiles for the 1cm source were comparable to that from 6MV in the order of 6mm for 10×10 cm{sup 2} field size at the depth of maximum dose. This could ascribe to Cobalt-60 beams producing lowerenergy secondary electrons. Although, the 2cm source have a larger penumbra however it could be still used for large targets with proportionally increased dose rate. For large lung targets, the difference between cobalt and 6MV plans is clinically insignificant. Our preliminary results showed that interchanging source sizes will allow cobalt beams for volumetric arc therapy of both small lesions and large tumors. Conclusion: The CybeRT system will be a cost effective machine capable of performing advanced radiation therapy treatments of both small tumors and large target volumes.

60 APPLIED LIFE SCIENCES↗

TH-AB-201-05: Determining the Direction Distribution of the Primary Radiation for a Cyberknife-M6

Purpose: Radiation protection regulatory differentiates between primary and scatter radiation. Whereas for conventional clinical linear accelerators the solid angle for primary radiation is planar, the Cyberknife (Accuray Inc., Sunnyvale, CA) may point its beam in all spatial directions. In order to be able to judge on radiation protection calculations for a Cyberknife-M6 vault, the direction distribution for delivered plans was evaluated based on clinical experiences. Methods: The log-files of 121 delivered patient treatment plans were exported, divided into cranial and extra-cranial treatments and the delivered monitor units (MU) together with the corresponding beam directions were analyzed. This MU-weighted spatial distribution was then projected to a 9.5 × 5.9 × 3.9 m{sup 3} vault, generating an “intensity map” using a binning of 50 × 50 cm{sup 2}. The factor of direction (FOD) is reported as a fraction of the total applied MUs to the walls, ceiling and floor in the perspective of a patient lying in head-first-supine position on the couch. In this study, the term intFOD refers to the integral FOD and maxFOD refers to the maximal FOD for a single bin. Results: For all kind of treatments and collimators, intFOD and maxFOD for the wall behind the patient’s head is 0.0. The intFOD for the floor varies between 0.65 and 0.74. For the ceiling, maxFOD is 0.002 and 0.0 for cranial and extra-cranial cases, respectively. The intFOD for the wall at the patient’s feet, is 0.094 for cranial and 0.005 for extra-cranial cases. There is nearly no difference between the maxFOD of the right and left wall for cranial cases, whereas for extra-cranial cases these numbers differ by a factor of 1.75. Conclusion: The varying FODs for the Cyberknife were quantified based on delivered treatment plans. These findings are helpful regarding the design of Cyberknife vaults.

60 APPLIED LIFE SCIENCES↗

Constituent Components of Out-of-Field Scatter Dose for 18-MV Intensity Modulated Radiation Therapy Versus 3-Dimensional Conformal Radiation Therapy: A Comparison With 6-MV and Implications for Carcinogenesis

Purpose: To characterize and compare the components of out-of-field dose for 18-MV intensity modulated radiation therapy (IMRT) versus 3-dimensional conformal radiation therapy (3D-CRT) and their 6-MV counterparts and consider implications for second cancer induction. Methods and Materials: Comparable plans for each technique/energy were delivered to a water phantom with a sloping wall; under full scatter conditions; with field edge abutting but outside the bath to prevent internal/phantom scatter; and with shielding below the linear accelerator head to attenuate head leakage. Neutron measurements were obtained from published studies. Results: Eighteen-megavolt IMRT produces 1.7 times more out-of-field scatter than 18-MV 3D-CRT. In absolute terms, however, differences are just approximately 0.1% of central axis dose. Eighteen-megavolt IMRT reduces internal/patient scatter by 13%, but collimator scatter (C) is 2.6 times greater than 18-MV 3D-CRT. Head leakage (L) is minimal. Increased out-of-field photon scatter from 18-MV IMRT carries out-of-field second cancer risks of approximately 0.2% over and above the 0.4% from 18-MV 3D-CRT. Greater photoneutron dose from 18-MV IMRT may result in further maximal, absolute increased risk to peripheral tissue of approximately 1.2% over 18-MV 3D-CRT. Out-of-field photon scatter remains comparable for the same modality irrespective of beam energy. Machine scatter (C+L) from 18 versus 6 MV is 1.2 times higher for IMRT and 1.8 times for 3D-CRT. It is 4 times higher for 6-MV IMRT versus 3D-CRT. Reduction in internal scatter with 18 MV versus 6 MV is 27% for 3D-CRT and 29% for IMRT. Compared with 6-MV 3D-CRT, 18-MV IMRT increases out-of-field second cancer risk by 0.2% from photons and adds 0.28-2.2% from neutrons. Conclusions: Out-of-field photon dose seems to be independent of beam energy for both techniques. Eighteen-megavolt IMRT increases out-of-field scatter 1.7-fold over 3D-CRT because of greater collimator scatter despite reducing internal/patient scatter. Out-of-field carcinogenic risk is thus increased (but improved in-field dose conformity may offset this). Potentially increased carcinogenic risk should be weighed against any benefit 18-MV IMRT may provide.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

SU-E-T-99: An Analysis of the Accuracy of TPS Extrapolation of Commissioning Data

Purpose: To investigate discrepancies between measured percent depth dose curves of a linear accelerator at depths beyond the commissioning data and those generated by the treatment planning system (TPS) via extrapolation. Methods: Relative depth doses were measured on an Elekta Synergy™ linac for photon beams of 6 -MV and 10-MV. SSDs for all curves were 100-cm and field sizes ranged from 4×4 to 35×35-cm{sup 2}. As most scanning tanks cannot provide depths greater than about 30-cm, percent depth dose measurements, extending 45-cm depths, were performed in Solid Water™ using a 0.125-cc ionization chamber (PTW model TN31012). The buildup regions of the curves were acquired with a parallel plate chamber (PTW model TN34001). Extrapolated curves were generated by the TPS (Phillips Pinnacle{sup 3} v. 9.6) by applying beams to CT images of 50-cm of Solid Water™ with density override set to 1.0-g/cc. Results: Percent difference between the two sets of curves (measured and TPS) was investigated. There is significant discrepancy in the buildup region to a depth of 7-mm. Beyond this depth, the two sets show good agreement. When analyzing the tail end of the curves, we saw percent difference of between 1.2% and 3.2%. The highest disagreement for the 6-MV curves was 10×10-cm{sup 2} (3%) and for the 10-MV curves it was the 35×35-cm{sup 2} (3.2%). Conclusion: A qualitative analysis of the measured data versus PDD curves generated by the TPS shows generally good agreement beyond 1-cm. However, a measurable percent difference was observed when comparing curves at depths beyond that provided by the commissioning data and at depths in the buildup region. Possible explanations for this include inaccuracies in modeling of the Solid Water™ or drift in beam energy since commissioning. Additionally, closer attention must be paid for measurements in the buildup region.

60 APPLIED LIFE SCIENCES↗

TU-FG-201-01: 18-Month Clinical Experience of a Linac Daily Quality Assurance (QA) Solution Using Only EPID and OBI

Purpose: To describe the clinical use of a Linear Accelerator (Linac) DailyQA system with only EPID and OBI. To assess the reliability over an 18-month period and improve the robustness of this system based on QA failure analysis. Methods: A DailyQA solution utilizing an in-house designed phantom, combined EPID and OBI image acquisitions, and a web-based data analysis and reporting system was commissioned and used in our clinic to measure geometric, dosimetry and imaging components of a Varian Truebeam Linac. During an 18-month period (335 working days), the Daily QA results, including the output constancy, beam flatness and symmetry, uniformity, TPR20/10, MV and KV imaging quality, were collected and analyzed. For output constancy measurement, an independent monthly QA system with an ionization chamber (IC) and annual/incidental TG51 measurements with ADCL IC were performed and cross-compared to Daily QA system. Thorough analyses were performed on the recorded QA failures to evaluate the machine performance, optimize the data analysis algorithm, adjust the tolerance setting and improve the training procedure to prevent future failures. Results: A clinical workflow including beam delivery, data analysis, QA report generation and physics approval was established and optimized to suit daily clinical operation. The output tests over the 335 working day period cross-correlated with the monthly QA system within 1.3% and TG51 results within 1%. QA passed with one attempt on 236 days out of 335 days. Based on the QA failures analysis, the Gamma criteria is revised from (1%, 1mm) to (2%, 1mm) considering both QA accuracy and efficiency. Data analysis algorithm is improved to handle multiple entries for a repeating test. Conclusion: We described our 18-month clinical experience on a novel DailyQA system using only EPID and OBI. The long term data presented demonstrated the system is suitable and reliable for Linac daily QA.

60 APPLIED LIFE SCIENCES↗