Development of Ta-based Superconducting Tunnel Junction X-ray Detectors for Fluorescence XAS
We are developing superconducting tunnel junction (STJ) soft X-ray detectors for chemical analysis of dilute samples by fluorescence-detected X-ray absorption spectroscopy (XAS). Our 36-pixel Nb-based STJ spectrometer covers a solid angle {Omega}/4{pi}{approx_equal}10{sup -3}, offers an energy resolution of {approx}10-20 eV FWHM for energies up to {approx}1 keV, and can be operated at total count rates of {approx}10{sup 6} counts/s. For increased quantum efficiency and cleaner response function, we have now started the development of Ta-based STJ detector arrays. Initial devices modeled after our Nb-based STJs have an energy resolution below 10 eV FWHM for X-ray energies below 1 keV, and pulse rise time discrimination can be used to improve their response function for energies up to several keV. We discuss the performance of the Ta-STJs and outline steps towards the next-generation of large STJ detector arrays with higher sensitivity.