An Examination of Causes of Multi-Wire Bias During SLD Instrumentation Testing in the IRT
The Icing Research Tunnel (IRT) at NASA Glenn Research Center has been using the SEA Multi-Element Sensor (aka, the multi-wire) as its primary liquid water content instrument since 2011. Recent tests in the IRT with four different multi-wire probes (i.e., different serial numbers) have indicated situations that can result in biasing of the probe data. The reference data used in this report are multi-wire data that have been proven to agree well with the icing blade in low-impingement rate conditions. Data from these recent tests suggest that the presence of a canister probe located 4.45 ft axially downstream and 2.5 ft to the side may have caused multi-wire data to read around 5% low at an airspeed of 150 kts. Additionally, two probes of different serial numbers were observed to measure 5-15% lower than the reference probes, and the causes have been linked to excessive solder on the sensing element and to an out of date element calibration. For these reasons, the authors suggest the community be aware of the possible aerodynamic influences on the probe inlet and outlet, as well as the importance of regular calibration and maintenance of the probes.