NASA NTRS1983
Bolometers operating at low temperatures (T greater than 0.3 K) are the best candidate for far infrared (lambda greater than or = 120 microns) and submillimeter detection from cold orbiting telescopes. The limits of sensitivity for such devices have been studied (Mather, 1983). Based on this work, we conclude that excellent sensitivity is possible given careful detector design. We have designed and are fabricating bolometers for operation at low temperatures. The crucial components in detector design are (1) heat capacity, (2) temperature sensing, and (3) absorption of infrared radiation. In this presentation, we discuss the choice of materials and fabrication techniques we have made in the interest of maximum sensitivity. In order that optimization of detector performance be reasonably easy to carry out, we have attempted to use conventional fabrication techniques wherever possible (photolithography, wire bonding). The use of such procedures allows us to make design changes on short time scales, which is essential in achieving optimal performance.