NASA NTRSDate not supplied
The XRISM/Resolve instrument cooling system uses adiabatic demagnetization refrigerators (ADRs) to cool the detectors to 50 mK and two types of cryocoolers to reduce heat load on the He tank, Joule-Thomson and Stirling cryocoolers. Resolve was designed with tunable cryocooler frequencies so that any mechanical interference and its effect on performance could be minimized. Measurements at different cryocooler drive frequencies were used to quantify the interference and narrow the acceptable range of frequency options. For Resolve the choice of cryocooler drive frequencies dramatically influences the instrument performance. Poor choices have been shown to degrade ADR cooling power and degrade the spectroscopic performance of the instrument. However, good choices have been shown to be free of undesirable features, allowing the instrument to satisfy its performance requirements with significant margin. Thus, the choice of drive frequencies is a critical factor for overall instrument performance. The drive frequencies of the Joule-Thomson cooler operating near 52 Hz, and the Stirling coolers operating near 15 Hz were adjusted in three distinct scan sequences to efficiently narrow the candidate frequency options. The frequency choices were based on stability of the 50 mK control thermometer, changes in the ADR demagnetization rate, and changes in detector noise. Scans across the drive frequency range of each cryocooler were performed multiple times during ground testing. Surprisingly, the results were not repeatable across test campaigns and did not conform to any discernable trend, but they were repeatable when remeasured within a single cool-down cycle. The conclusion was that the good/bad drive frequencies changed during Dewar warmup/cooldown, and possibly after mechanical vibration tests. Therefore, the settings could not be finalized before launch, and scans during instrument commissioning were necessary. This paper describes the cryocooler frequency scan measurements and compares the results during instrument commissioning to those performed during ground tests.