Reproducibility of Global and Local Reconstruction of Three-Dimensional Micro-Computed Tomography of Iliac Crest Biopsies
Variation in computed tomography (CT) image grayscale and spatial geometry due to specimen orientation, magnification, voxel size, differences in X-ray photon energy and limited field-of-view during the scan, were evaluated in repeated micro-CT scans of iliac crest biopsies and test phantoms. Using the micro-CT scanner on beamline X2B at the Brookhaven National Laboratory's National Synchrotron Light Source, 3-D micro-CT images were generated. They consisted of up to 1024 X 24002, 4-mum cubic voxels, each with 16-bit gray-scale. We also reconstructed the images at 16-, 32-, and 48-mum voxel resolution. Scan data were reconstructed from the complete profiles using filtered back-projection and from truncated profiles using profile-extension and with a Local reconstruction algorithm. Three biopsies and one bonelike test phantom were each rescanned at three different times at annual intervals. For the full-data-set reconstructions, the reproducibility of the estimates of mineral content of bone at mean bone opacity value, was {+-}28.8 mg/cm3, i.e., 2.56%, in a 4-mum cubic voxel at the 95% confidence level. The reproducibility decreased with increased voxel size. The interscan difference in imaged bone volume ranged from 0.86 {+-} 0.64% at 4-mum voxel resolution, and 2.64 {+-} 2.48% at 48 mum.