Signal detection in l/f noise of SQUID magnetometers
It is shown that the variance on the SQUID power spectrum in the l/f low frequency region is well behaved, i.e., any small frequency band may be treated as white noise in standard power spectrum estimation theory. Specifically a calibration signal is examined at 0.017 Hz with an equivalent energy referred to the SQUID input coil of 1 times 10 to the -30th J and a digitally recorded and analyzed record of 140 hr duration obtained an optimum S/N better than 400. The results are in good agreement with theory. In addition no deviation from the l/f dependence of the noise energy spectrum is seen down to frequencies below 10 to the -5th Hz. A commercially available SQUID and electronics system were used.