Counterflow Virtual Impactor (CVI) Inlet Aboard Aircraft (INLETCVI-AIR) Instrument Handbook
The counterflow virtual impactor (CVI) inlet separates out and samples cloud elements via inertial impaction. Warm, dry, particle-free carrier gas of known composition is pumped to the tip of the inlet in the direction opposite of the flow of air. The flow lines of the incoming free stream air are split. Larger droplets and ice crystals suspended in the sample air have enough inertia to penetrate the counterflow and enter the sample flow, while smaller, unactivated particles follow the streamlines of the air flowing around the inlet’s tip (Noone et al. 1988). The instrument is capable of sampling cloud elements as a function of cut size by changing the velocity of counterflow at the probe tip. The CVI cut size is usually defined as a diameter of a smallest particle for which penetration rate is 50%. For the U.S. Department of Energy (DOE) Atmospheric Radiation Measurement (ARM) user facilityGulfstream-159 aircraft’s CVI inlet, that value is usually around 13 microns. The dry heated air, normally used as the carrier gas, also serves to evaporate water from the hydrometeor, leaving the residuals to be sampled by in-cabin aerosol instrumentation. The number concentration of the droplet residue in the CVI sample line is usually increased with respect to corresponding cloud element number concentration. The CVI performance as a concentrator is usually described via CVI enhancement factor, which is a conversion factor between ambient volumetric concentration (not adjusted to standard temperature and pressure) and volumetric concentration in the CVI sample line.