Calibration of diode laser spectra using a confocal etalon
The dual-beam diode laser spectrometer described by Jennings (1980) is adapted to use a 50-cm confocal etalon for frequency calibration. The collimated radiation from the laser is split at a wedged ZnSe window, and the reference beam is then focused at the midpoint of the etalon length. After the etalon, the reference beam is recollimated and continues its regular path to the monochromator and detectors. An aperture is placed before the etalon in order to limit the entrance beam diameter to approximately 5 mm. Both ends of the etalon are furnished with two-axis adjustments. Initial alignment is achieved using an He-Ne laser, and final optimization involves adjustment of the cavity length as well as the etalon pitch and yaw. The 50-cm confocal etalon produces fringes separated by 150 MHz (0.005/cm). With the aid of a CO2 laser, it is found to have fringe widths (FWHM) of 2 MHz. The confocal etalon makes it possible to improve the accuracy of relative frequency measurements in diode laser spectra and to check the spectral purity and stability of the laser during the recording of spectra.