Compact TDLAS based optical sensor for ppb-level ethane detection by use of a 3.34 μm room-temperature CW interband cascade laser
A mid-infrared ethane (C 2 H 6 ) sensor based on a wavelength modulation spectroscopy (WMS) technique was developed using a thermoelectrically cooled (TEC), continuous-wave (CW) interband cascade laser (ICL) emitting at 3.34 μm and a dense multi-pass gas cell (MPGC, 17 × 6.5 × 5.5 cm 3 ) with a 54.6 m optical path length. A compact optical sensor system with a physical size of 35.5 × 18 × 12.5 cm 3 was designed and constructed. An ICL was employed for targeting a strong C 2 H 6 line at 2996.88 cm -1 at <100 Torr gas pressure in the fundamental absorption band of C 2 H 6 . The sensor performance, including the minimum detection limit (MDL) and the stability were improved by reducing the effect of laser power drift by means of the 2f/1f-WMS technique. A MDL of ~1.2 parts per billion (ppbv) for 2f-WMS and ~1.0 ppbv for 2f/1f-WMS were achieved, respectively, with a measurement time of 4 s. The MDL was further improved from 299 pptv (@108 s for 2f-WMS) to 239 pptv (@208 s for 2f/1f-WMS), based on an Allan deviation analysis.The rise time (@0 → 100 ppbv) and fall time (@100 → 0 ppbv) were determined to be ~64 s and ~48 s,respectively, at a gas pressure of <100 Torr for the C 2 H 6 sensor operation.