SU-E-T-182: Clinical Implementation of TG71-Based Electron MU Calculation and Comparison with a Commercial Secondary Calculation
Purpose: The TG-71 report was published in 2014 to present standardized methodologies for MU calculations and determination of dosimetric quantities. This work explores the clinical implementation of a TG71-based electron MU calculation algorithm and compares it with a recently released commercial secondary calculation program–Mobius3D (Mobius Medical System, LP). Methods: TG-71 electron dosimetry data were acquired, and MU calculations were performed based on the recently published TG-71 report. The formalism in the report for extended SSD using air-gap corrections was used. The dosimetric quantities, such PDD, output factor, and f-air factors were incorporated into an organized databook that facilitates data access and subsequent computation. The Mobius3D program utilizes a pencil beam redefinition algorithm. To verify the accuracy of calculations, five customized rectangular cutouts of different sizes–6×12, 4×12, 6×8, 4×8, 3×6 cm{sup 2}–were made. Calculations were compared to each other and to point dose measurements for electron beams of energy 6, 9, 12, 16, 20 MeV. Each calculation / measurement point was at the depth of maximum dose for each cutout in a 10×10 cm{sup 2} or 15×15cm{sup 2} applicator with SSDs 100cm and 110cm. Validation measurements were made with a CC04 ion chamber in a solid water phantom for electron beams of energy 9 and 16 MeV. Results: Differences between the TG-71 and the commercial system relative to measurements were within 3% for most combinations of electron energy, cutout size, and SSD. A 5.6% difference between the two calculation methods was found only for the 6MeV electron beam with 3×6 cm{sup 2}cutout in the 10×10{sup 2}cm applicator at 110cm SSD. Both the TG-71 and the commercial calculations show good consistency with chamber measurements: for 5 cutouts, <1% difference for 100cm SSD, and 0.5–2.7% for 110cm SSD. Conclusions: Based on comparisons with measurements, a TG71-based computation method and a Mobius3D program produce reasonably accurate MU calculations for electron-beam therapy.