Experimental system for the control of surgically induced infections
The results are presented of the development tests performed on the experimental system for the control of surgically induced infections. Tests were performed on the portable clean room to demonstrate assembly, collapsability, portability and storage. Collapsing, relocating and storing within the surgery room can be accomplished in 12 minutes. The storage envelope dimensions are 1.64 m x 4.24 m x 2.62 m high. The disassembly transfer to another room, and reassembly were demonstrated. The laminar air flow velocity profile within the enclosure was measured. In the undisturbed area of the enclosure the air flow met the Federal Standard 209a requirements of 27.45 meters per minute + or - 6.10 meters per minute. Smoke tests with simulated surgery equipment and personnel in the enclosure did not indicate any detrimental air flow patterns. It is concluded that the system as designed will perform the functions required for its intended use.