Impulse response measurements of fast scintillator-based current mode detectors
In this study, experimental measurements were performed to characterize the impulse response of seven different gamma ray detector prototypes being considered to be fielded as a current mode diagnostic for neutron-diagnosed subcritical experiments at the Nevada National Security Site (NNSS). For such experiments, the impulse response of the fielded detector is required to have a full-width half-maximum (FWHM) of approximately 5 ns or less and a low amplitude tail. Each of the detector prototypes evaluated in this work were selected based on the potential to meet this performance requirement. An impulse of bremsstrahlung x-rays created from 2 MeV electrons from the NNSS’s Transformational Diagnostics and Imaging, Los Alamos, linear accelerator was measured with each detector. The measured impulse response was evaluated for each detector configuration. A comparison of these impulse responses revealed a relationship between fast timing performance and scintillator material, geometry, surface reflectivity conditions, and photomultiplier tube (PMT) selection. The detector configuration yielding an impulse response with the shortest FWHM of 3.82 ± 0.11 ns and possessing a low amplitude tail was a 5-inch (12.7-cm) diameter, 5-inch (12.7-cm) height cylindrical EJ-399-17-VI liquid scintillator in a black-painted housing coupled to an Adit D798B 5-inch (12.7-cm) PMT by means of a 0.079-inch (0.2-cm) EJ-560 optical coupling pad.