Kilopixel X-Ray Microcalorimeter Arrays for Astrophysics: Device Performance and Uniformity

We are developing kilo-pixel arrays of TES microcalorimeters to enable high-resolution X-ray imaging spectrometers for future X-ray observatories and laboratory astrophysics experiments. Our current array design was targeted as a prototype for the X-ray Microcalorimeter Spectrometer proposed for the International X-ray Observatory, which calls for a 40x40-pixel core array of 300 micron devices with 2.5 e V energy resolution (at 6 keV). Here we present device characterization of our 32x32 arrays, including X-ray spectral performance of individual pixels within the array. We present our results in light of the understanding that our Mo/Au TESs act as weak superconducting links, causing the TES critical current (Ic) and transition shape to oscillate with applied magnetic field (B). We show Ic(B) measurements and discuss the uniformity of these measurements across the array, as well as implications regarding the uniformity of device noise and response. In addition, we are working to reduce pixel-to-pixel electrical and thermal crosstalk; we present recent test results from an array that has microstrip wiring and an angle-evaporated Cu backside heatsinking layer, which provides Cu coverage on the four sidewalls of the silicon wells beneath each pixel.

Eckart, M. E.↗

Fabrication of a Kilopixel Array of Superconducting Microcalorimeters with Microstripline Wiring

A document describes the fabrication of a two-dimensional microcalorimeter array that uses microstrip wiring and integrated heat sinking to enable use of high-performance pixel designs at kilopixel scales (32 X 32). Each pixel is the high-resolution design employed in small-array test devices, which consist of a Mo/Au TES (transition edge sensor) on a silicon nitride membrane and an electroplated Bi/Au absorber. The pixel pitch within the array is 300 microns, where absorbers 290 microns on a side are cantilevered over a silicon support grid with 100-micron-wide beams. The high-density wiring and heat sinking are both carried by the silicon beams to the edge of the array. All pixels are wired out to the array edge. ECR (electron cyclotron resonance) oxide underlayer is deposited underneath the sensor layer. The sensor (TES) layer consists of a superconducting underlayer and a normal metal top layer. If the sensor is deposited at high temperature, the ECR oxide can be vacuum annealed to improve film smoothness and etch characteristics. This process is designed to recover high-resolution, single-pixel x-ray microcalorimeter performance within arrays of arbitrarily large format. The critical current limiting parts of the circuit are designed to have simple interfaces that can be independently verified. The lead-to-TES interface is entirely determined in a single layer that has multiple points of interface to maximize critical current. The lead rails that overlap the TES sensor element contact both the superconducting underlayer and the TES normal metal

Chervenak, James↗

Development of a Real-Time Pulse Processing Algorithm for TES-Based X-Ray Microcalorimeters

We report here a real-time pulse processing algorithm for superconducting transition-edge sensor (TES) based x-ray microcalorimeters. TES-based. microca1orimeters offer ultra-high energy resolutions, but the small volume of each pixel requires that large arrays of identical microcalorimeter pixe1s be built to achieve sufficient detection efficiency. That in turn requires as much pulse processing as possible must be performed at the front end of readout electronics to avoid transferring large amounts of data to a host computer for post-processing. Therefore, a real-time pulse processing algorithm that not only can be implemented in the readout electronics but also achieve satisfactory energy resolutions is desired. We have developed an algorithm that can be easily implemented. in hardware. We then tested the algorithm offline using several data sets acquired with an 8 x 8 Goddard TES x-ray calorimeter array and 2x16 NIST time-division SQUID multiplexer. We obtained an average energy resolution of close to 3.0 eV at 6 keV for the multiplexed pixels while preserving over 99% of the events in the data sets.

Tan, Hui↗

Study of the Dependency on Magnetic Field and Bias Voltage of an AC-Biased TES Microcalorimeter

At SRON we are studying the performance of a Goddard Space Flight Center single pixel TES microcalorimeter operated in an AC bias configuration. For x-ray photons at 6 keV the pixel shows an x-ray energy resolution Delta E(sub FWHM) = 3.7 eV, which is about a factor 2 worse than the energy resolution observed in an identical DC-biased pixel. In order to better understand the reasons for this discrepancy we characterized the detector as a function of temperature, bias working point and applied perpendicular magnetic field. A strong periodic dependency of the detector noise on the TES AC bias voltage is measured. We discuss the results in the framework of the recently observed weak-link behaviour of a TES microcalorimeter.

Gottardi, L.↗

Designing the X-Ray Microcalorimeter Spectrometer for Optimal Science Return

Recent advances in X-ray microcalorimeters enable a wide range of possible focal plane designs for the X-ray Microcalorimeter Spectrometer (XMS) instrument on the future Advanced X-ray Spectroscopic Imaging Observatory (AXSIO) or X-ray Astrophysics Probe (XAP). Small pixel designs (75 microns) oversample a 5-10" PSF by a factor of 3-6 for a 10 m focal length, enabling observations at both high count rates and high energy resolution. Pixel designs utilizing multiple absorbers attached to single transition-edge sensors can extend the focal plane to cover a significantly larger field of view, albeit at a cost in maximum count rate and energy resolution. Optimizing the science return for a given cost and/or complexity is therefore a non-trivial calculation that includes consideration of issues such as the mission science drivers, likely targets, mirror size, and observing efficiency. We present a range of possible designs taking these factors into account and their impacts on the science return of future large effective-area X-ray spectroscopic missions.

X-rays↗

Kilopixel X-Ray Microcalorimeter Arrays for Astrophysics: Device Performance and Uniformity

We are developing kilopixel arrays of TES microcalorimeters to enable high-resolution x-ray imaging spectrometers for future x-ray observatories and laboratory astrophysics experiments. Our current array design was targeted as a prototype for the X-ray Microcalorimeter Spectrometer proposed for the International X-ray Observatory, which calls for a 40×40-pixel core array of 300 micron devices with 2.5 eV energy resolution (at 6 keV). Here we present device characterization of our 32×32 arrays, including x-ray spectral performance of individual pixels within the array. We present our results in light of the understanding that our Mo/Au TESs act as weak superconducting links, causing the TES critical current (I(sub c)) and transition shape to oscillate with applied magnetic field (B). We show I(sub c)(B) measurements and discuss the uniformity of these measurements across the array, as well as implications regarding the uniformity of device noise and response. In addition, we are working to reduce pixel-to-pixel electrical and thermal crosstalk; we present recent test results from an array that has microstrip wiring and an angle-evaporated copper backside heatsinking layer, which provides copper coverage on the four sidewalls of the silicon wells beneath each pixel.

Transition-edge sensors A[center dot] Superconduct↗

Calibration of the Microcalorimeter Spectrometer On-Board the Hitomi (Astro-H) Observatory (invited)

The Hitomi Soft X-ray Spectrometer (SXS) was a pioneering non-dispersive imaging x-ray spectrometer with 5 eV FWHM energy resolution, consisting of an array of 36 silicon-thermistor microcalorimeters at the focus of a high-throughput soft x-ray telescope. The instrument enabled astrophysical plasma diagnostics in the 0.3-12 keV band. We introduce the SXS calibration strategy and corresponding ground calibration measurements that took place from 2012-2015, including both the characterization of the microcalorimeter array and measurements of the x-ray transmission of optical blocking filters.

ASTRO H↗

Toward Large FOV High-Resolution X-Ray Imaging Spectrometer: Microwave Multiplexed Readout of 32 TES Microcalorimeters

We performed a small-scale demonstration at GSFC of high-resolution x-ray TES microcalorimeters read out using a microwave SQUID multiplexer. This work is part of our effort to develop detector and readout technologies for future space based x-ray instruments such as the microcalorimeter spectrometer envisaged for Lynx, a large mission concept under development for the Astro 2020 Decadal Survey. In this paper we describe our experiment, including details of a recently designed, microwave-optimized low-temperature setup that is thermally anchored to the 50 mK stage of our laboratory ADR. Using a ROACH2 FPGA at room temperature, we simultaneously read out 32 pixels of a GSFC-built detector array via a NIST-built multiplexer chip with Nb coplanar waveguide resonators coupled to RF SQUIDs. The resonators are spaced 6 MHz apart (at approx. 5.9 GHz) and have quality factors of approximately 15,000. Using flux-ramp modulation frequencies of 160 kHz we have achieved spectral resolutions of 3 eV FWHM on each pixel at 6 keV. We will present the measured system-level noise and maximum slew rates, and briefly describe the implications for future detector and readout design.

Yoon, Wonsik↗

Toward Large Field-Of-View High-Resolution X-Ray Imaging Spectrometers: Microwave Multiplexed Readout of 28 TES Microcalorimeters

We performed small-scale demonstrations at GSFC of high-resolution Xray TES microcalorimeters read out using a microwave SQUID multiplexer. This work is part of our effort to develop detector and readout technologies for future space-based X-ray instruments such as the microcalorimeter spectrometer envisaged for Lynx, a large mission concept under development for the Astro 2020 Decadal Survey. In this paper we describe our experiment, including details of a recently designed, microwave-optimized low-temperature setup that is thermally anchored to the 55mKstage of our laboratory ADR. Using aROACH2 FPGA at room temperature, we read out pixels of a GSFC-built detector array via a NIST-built multiplexer chip with Nb coplanar waveguide resonators coupled to rf-SQUIDs. The resonators are spaced 6 MHz apart (at ∼ 5.9 GHz) and have quality factors of ∼ 15,000. In our initial demonstration, we used flux-ramp modulation frequencies of 125 kHz to read out 5 pixels simultaneously and achieved spectral resolutions of 2.8-3.1 eV FWHM at 5.9 keV. Our subsequent work is ongoing: to-date we have achieved a median spectral resolution of 3.4 eV FWHM at 5.9 keV while reading out 28 pixels simultaneously with flux-ramp frequencies of 160 kHz. We present the measured system-level noise and maximum slew rates and briefly describe our future development work.

Yoon, W.↗

Thermal Fluctuation Noise in Mo/Au Superconducting Transition-Edge Sensor Microcalorimeters

In many superconducting transition-edge sensor (TES) microcalorimeters, the measured electrical noise exceeds theoretical estimates based on a thermal model of a single body thermally connected to a heat bath. Here, we report on noise and complex impedance measurements of a range of designs of TESs made with a Mo/Au bilayer. We have fitted the measured data using a two-body model, where the x-ray absorber and the TES are connected by an internal thermal conductance Gae. We find that the so-called excess noise measured in these devices is consistent with the noise generated from the internal thermal fluctuations between the x-ray absorber and the TES. Our fitted parameters are consistent with the origin of Gae being from the finite thermal conductance of the TES itself. These results suggest that even in these relatively low resistance Mo/Au TESs, the internal thermal conductance of the TES may add significant additional noise and could account for all the measured excess noise. Furthermore, we find that around regions of the superconducting transition with rapidly changing derivative of resistance with respect to temperature, an additional noise mechanism may dominate. These observations may lead to a greater understanding of TES devices and allow the design of TES microcalorimeters with improved performance.

Wakeham, N. A.↗

Microwave SQUID Multiplexing for the Lynx X-Ray Microcalorimeter

The Lynx x-ray microcalorimeter (LXM) is an imaging spectrometer for the Lynx satellite mission, an x-ray telescope being considered by NASA to be a new flagship mission. Lynx will enable unique astrophysical observations into the x-ray universe due to its high angular resolution and large field of view. The LXM consists of an array of over 100,000 pixels and poses a significant technological challenge to achieve the high degree of multiplexing required to read out these sensors. We discuss the details of microwave superconducting quantum interference device (SQUID) multiplexing and describe why it is ideally suited to the needs of the LXM. This case is made by summarizing the current and predicted performance of microwave SQUID multiplexing and describing the steps needed to optimize designs for all the LXM arrays. Finally, we describe our plan to advance the technology readiness level (TRL) of microwave SQUID multiplexing of the LXM microcalorimeters to TRL-5 by 2024.

Bennett, Douglas A.↗

Thermal crosstalk measurements and simulations for an X-ray microcalorimeter array

Arrays of high-density microcalorimeters require careful heat sinking in order to minimize the thermal crosstalk between nearby pixels. For the array of microcalorimeters developed for the Athena X-ray Integral Field Unit instrument, which has more than 3000 pixels on a 275 µm pitch, it is essential to address this problem in order to meet the energy-resolution requirements. The instrument’s energy-resolution budget requires that the impact of the thermal crosstalk on the energy resolution be a contribution that, added in quadrature to other energy-resolution contributions, is less than 0.2 eV. This value results in a derived requirement that the ratio between the amplitude of the crosstalk signal to an X-ray pulse (for example at 6 keV) is less than 1 × 10(exp −3) (for the first neighbor), less than 4 × 10(exp −4) (for the diagonal neighbor) and less than 8 × 10(exp −5) (for the second nearest neighbor). We have measured the thermal crosstalk levels between pixels in various geometries and configurations. The results show a crosstalk ratio which is at least a factor of 4 lower than the derived requirement. We also developed a finite element (FEM) 2D thermal model to predict the thermal behavior of large-scale arrays. This model successfully simulates the measured data in terms of pulse amplitude and time constants.

Antoine R. Miniussi↗

Lynx x-ray microcalorimeter

Lynx is an x-ray telescope, one of four large satellite mission concepts currently being studied by NASA to be a flagship mission. One of Lynx’s three instruments is an imaging spectrometer called the Lynx x-ray microcalorimeter (LXM), an x-ray microcalorimeter behind an x-ray optic with an angular resolution of 0.5 arc sec and ∼2 sq. m of area at 1 keV. The LXM will provide unparalleled diagnostics of distant extended structures and, in particular, will allow the detailed study of the role of cosmic feedback in the evolution of the Universe. We discuss the baseline design of LXM and some parallel approaches for some of the key technologies. The baseline sensor technology uses transition-edge sensors, but we also consider an alternative approach using metallic magnetic calorimeters. We discuss the requirements for the instrument, the pixel layout, and the baseline readout design, which uses microwave superconducting quantum interference devices and high-electron mobility transistor amplifiers and the cryogenic cooling requirements and strategy for meeting these requirements. For each of these technologies, we discuss the current technology readiness level and our strategy for advancing them to be ready for flight. We also describe the current system design, including the block diagram, and our estimate for the mass, power, and data rate of the instrument.

Simon R. Bandler↗

Calibration and Testing of Small High-Resolution Transition Edge Sensor Microcalorimeters With Optical Photons

Pulses of narrow line-width optical photons can be used to calibrate and test sub-2 eV full-width at half-maximum (FWHM) energy resolution transition-edge sensor (TES) microcalorimeters at low energies (<1 keV),where it is very challenging to obtain X-ray calibration lines comparable to (or narrower than) the detector resolution. This scheme depends on the ability to resolve the number of 3 eV photons in each pulse, which we have recently demonstrated up to photon numbers of about 300. AtLTD-18 we showed preliminary results obtained with this technique on a 0.25 eV baseline resolution TES microcalorimeter designed for the ultra-high-resolution sub-array of the Lynx mission. The line-shape was well described by a simple Gaussian. However, the difficulty of delivering photons to the small 46 μm square absorbers resulted in a large thermal crosstalk signal, whose random nature is expected to rapidly degrade the observed energy resolution towards higher photon numbers/energies. We have since improved the coupling between the optical fiber and the TES absorber and report here our current results.

Felix T Jaeckel↗

Development of the Microcalorimeter Array for the Line Emission Mapper (LEM) X-Ray Probe

The Line Emission Mapper (LEM) is a proposed NASA probe-class mission designed to study the formation of structure in the Universe. LEM will have a 14k pixel array of transition-edge sensor (TES) microcalorimeters that provide a spectral resolution of < 2.5 eV and a bandpass of 0.2-2 keV. In this paper, we report on the status and plan for continued development of the LMS microcalorimeter. This includes details of the TES and x-ray absorber design, and the measured performance in prototype arrays. We present a detailed budget for the energy resolution and show how the properties of the LEM pixels are well optimized for the baseline time-division-multiplexing (TDM) that will be used read out the array. Finally, we describe the design and performance of the first full scale LMS arrays which include the necessary mechanical and electrical interfaces needed for LEM and are a precursor to an engineering model detector.

Stephen J Smith↗

Development of Transition-Edge Sensor X-ray Microcalorimeter Linear Array for Compton Scattering and Energy Dispersive Diffraction Imaging

In this paper we present a strip transition-edge sensor microcalorimeter linear array detector developed for energy dispersive X-ray diffraction imaging and Compton scattering applications. The prototype detector is an array of 20 transition-edge sensors with absorbers in strip geometry arranged in a linear array. We discuss the fabrication steps needed to develop this array including Mo/Cu bilayer, Au electroplating, and proofof-principle fabrication of long strips of SiNx membranes. We demonstrate minimal unwanted effect of strip geometry on X-ray pulse response and showlinear relationship of 1/pulse height and pulse decay times with absorber length. For the absorber lengths studied, our preliminary measurements show energy resolutions of 40-180eV near 17 keV. Furthermore, we show that the heat flow to the cold bath is nearly independent of the absorber area and depends on the SiNx membrane geometry.

47 OTHER INSTRUMENTATION↗

Automated co-adding and energy calibration of large array microcalorimeter data with zero sample knowledge

State-of-the-art microcalorimeter spectrometers now contain large detector arrays with hundreds of individual pixels. Each individual pixel outputs a unique and non-linear response with respect to deposited energy. This work describes a pattern-recognition algorithm to combine these responses into a single energy-calibrated histogram, referred to as co-adding pixels. Photo-peaks from different pixels are matched together based upon how well the match aligns the centroids and heights of neighboring peaks. This usually results in around 100 co-adding calibration points from 30 to 300 keV for a several day acquisition of plutonium items with masses between 0.5 and 10 grams. An additional algorithm energy-calibrates this co-added spectrum using the fluoresced K x-ray emissions from a tantalum absorber and inherent x-ray escape peaks from the tin absorbers. Both algorithms operate without knowledge of the source and are fully automated. This work presents results from the acquisitions of high and low burnup plutonium, 10% enriched uranium, a 153 Gd calibration source, and a 57 Co+ 166m Ho calibration source. In all measurements, resolution defined as the full-width at half-maximum (FWHM) of photo-peaks is preserved between the individual pixel and co-added spectra at around 65 eV for incident photon energies between 60 and 208 keV. The energy calibration algorithm is approximate and yields a calibration curve off by an average of around 200 eV for incident photon energies between 60 and 208 keV.

47 OTHER INSTRUMENTATION↗

Improved plutonium and americium photon branching ratios from microcalorimeter gamma spectroscopy

Photon branching ratios are critical input data for activities such as nuclear materials protection and accounting because they allow material compositions to be extracted from measurements of gamma-ray intensities. Uncertainties in these branching ratios are often a limiting source of uncertainty in composition determination. Here, we use high statistics, high resolution - (60–70 eV full-width-at-half-maximum at 100 keV) gamma-ray spectra acquired using microcalorimeter sensors to substantially reduce the uncertainties for 11 plutonium ( 238 Pu, 239 Pu, 241 Pu) and 241 Am branching ratios important for material control and accountability and nuclear forensics in the energy range of 125 keV to 208 keV. Lastly, we show a reduction in uncertainty of over a factor of three for one branching ratio and a factor of 2–3 for four branching ratios.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗