Calibration of a compact ASIC-based data acquisition system for neutron/$γ$ discrimination and spectroscopy with organic scintillators
Segmented neutron detectors that use silicon photomultipliers (SiPMs) are receiving significant attention in nuclear security applications. Some of these detectors employ hundreds of channels and would therefore benefit from the use of high-channel-density and low-cost-per-channel data acquisition (DAQ) systems. Candidate DAQ systems that meet these requirements exist, but few perform full waveform digitization, which permits neutron and gamma-ray interaction discrimination via pulse shape. In this work, we study the performance of the TOFPET2 (PETsys Electronics), an ASIC-based DAQ designed for positron emission tomography, which has been adapted to provide sensitivity to pulse shape by the use of variable-period charge integration. We use a light-emitting diode to calibrate a combination of an ON Semiconductor (SensL) 60035-64P J-Series SiPM and TOPFET2 DAQ and evaluate the linearity of their response and dynamic range. Here, the calibration curve was obtained by comparing the DAQ response to that measured with a photodiode interfaced with a traditional waveform digitizer. This calibration process was used for rudimentary spectroscopy of various neutron and sources.