Poster - 28: Shielding of X-ray Rooms in Ontario in the Absence of Best Practice
This poster will be strictly based on the Healing Arts Radiation Protection (HARP) Act, Regulation 543 under this Act (X-ray Safety Code), and personal communication the presenting author has had. In Ontario, the process of approval of an X-ray machine installation by the Director of the X-ray Inspection Service (XRIS) follows a certain protocol. Initially, the applicant submits a series of forms, including recommended shielding amounts, in order to satisfy the law. This documentation is then transferred to a third-party vendor (i.e. a professional engineer – P.Eng.) outsourced by the Ministry of Health and Long-term Care (MOHLTC). The P.Eng. then evaluates the submitted documentation for appropriate fulfillment of the HARP Act and Reg. 543 requirements. If the P.Eng.’s evaluation of the documentation is to their satisfaction, the XRIS is then notified. Finally, the Director will then issue a letter of approval to install the equipment at the facility. The methodology required to be used by the P.Eng. in order to determine the required amounts of protective barriers, and recommended to be used by the applicant, is contained within Safety Code 20A. However, Safety Code 35 has replaced the obsolete Safety Code 20A document and employs best practices in shielding design. This talk will focus further on specific intentions and limitations of Safety Code 20A. Furthermore, this talk will discuss the definition of the “practice of professional engineering” in Ontario. COMP members who are involved in shielding design are strongly encouraged to attend.