Optimizing the spatial resolution and gamma discrimination of SiPM-based Anger cameras

SiPM Anger cameras have been designed for use as 2-dimensional thermal neutron detectors for scientific applications such as at single crystal diffraction instruments. These cameras utilize a neutron sensitive scintillator and an array of SiPM photosensors, separated by a light spreading glass. While this optics package is very effective at detecting and positioning neutrons, it is also sensitive to gamma rays, which are a source of unwanted noise. In this work, we underwent a search of scintillator and spreader glass thicknesses to determine the optimal combination for maximizing spatial resolution. Both experimental results and GEANT4 simulations are presented. Further, to address the issue of gamma ray detection, we have also designed and presented a multi-layer scintillator that reduces the amount of scintillation light produced via gamma-ray energy deposition, allowing for easier pulse-height discrimination. A sample of this geometry has been produced that results in a factor of 2 improvement over an equivalent thickness monolithic scintillator.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

A Quench Detection and Monitoring System for Superconducting Magnets at Fermilab

A quench detection system was developed for protecting and monitoring the superconducting solenoids for the Muon-to-Electron Conversion Experiment (Mu2e) at Fermilab. The quench system was designed for a high level of dependability and long-term continuous operation. It is based on three tiers: Tier-I, FPGA-based Digital Quench Detection (DQD); Tier-II, Analog Quench Detection (AQD); and Tier-3, the quench controls and data management system. The Tier-I and Tier-II are completely independent and fully redundant systems. The Tier-3 system is based on National Instruments (NI) C-RIO and provides the user interface for quench controls and data management. It is independent from Tiers I & II. The DQD provides both quench detection and quench characterization (monitoring) capability. Both DQD and AQD have built-in high voltage isolation and user programmable gains and attenuations. The DQD and AQD also includes user configured current dependent thresholding and validation times. A 1 st article of the three-tier system was fully implemented on the new Fermilab magnet test stand for the HL-LHC Accelerator Up-grade Project (AUP). It successfully provided quench protection and monitoring (QPM) for a cold superconducting bus test in November 2020. The Mu2e quench detection design has since been implemented for production testing of the AUP magnets. A detailed description of the system along with results from the AUP superconducting bus test will be presented.

monitoring↗

A Graph Neural Network Surrogate Model for hls4ml

Recent advancements in use of machine learning (ML) techniques on field-programmable gate arrays (FPGAs) have allowed for the implementation of embedded neural networks with extremely low latency. This is invaluable for particle detectors at the Large Hadron Collider, where latency and used area are strictly bounded. The hls4ml framework is a procedure that converts trained ML model software to a synthesis result to can be used on an FPGA. However, running the pipeline is a time-consuming procedure, and there is a strong risk of failure. In particular, it may not be possible to successfully convert a model into a synthesis result, or the resource consumption of the model may exceed the resources of the target FPGA. To aid with this development, we introduce wa-hls4ml, a surrogate model using a graph neural network to emulate the structure of the source models. The goal is to estimate the chance of success and resource consumption of a given model when passed through the hls4ml pipeline, without needing to run the pipeline.

Plotnikov, Dennis↗

The energy spectrum of cosmic rays beyond the turn-down around 10 17 eV as measured with the surface detector of the Pierre Auger Observatory

We present a measurement of the cosmic-ray spectrum above 100 PeV using the part of the surface detector of the Pierre Auger Observatory that has a spacing of 750 m. An inflection of the spectrum is observed, confirming the presence of the so-called second-knee feature. The spectrum is then combined with that of the 1500 m array to produce a single measurement of the flux, linking this spectral feature with the three additional breaks at the highest energies. The combined spectrum, with an energy scale set calorimetrically via fluorescence telescopes and using a single detector type, results in the most statistically and systematically precise measurement of spectral breaks yet obtained. These measurements are critical for furthering our understanding of the highest energy cosmic rays.

79 ASTRONOMY AND ASTROPHYSICS↗

Synchronous detection of cosmic rays and correlated errors in superconducting qubit arrays

Quantum information processing at scale will require sufficiently stable and long-lived qubits, likely enabled by error-correction codes. Several recent superconducting-qubit experiments, however, reported observing intermittent spatiotemporally correlated errors that would be problematic for conventional codes, with ionizing radiation being a likely cause. Here, we directly measured the cosmic-ray contribution to spatiotemporally correlated qubit errors. We accomplished this by synchronously monitoring cosmic-ray detectors and qubit energy-relaxation dynamics of 10 transmon qubits distributed across a 5 × 5 × 0.35 mm 3 silicon chip. Cosmic rays caused correlated errors at a rate of $1/\left(592\begin{array}{c}+48\\ -41\end{array}\,{\rm{s}}\right)$, accounting for 17.1 ± 1.3% of all such events. Our qubits responded to essentially all of the cosmic rays and their secondary particles incident on the chip, consistent with the independently measured arrival flux. Moreover, we observed that the landscape of the superconducting gap in proximity to the Josephson junctions dramatically impacts the qubit response to cosmic rays. Given the practical difficulties associated with shielding cosmic rays, our results indicate the importance of radiation hardening—for example, superconducting gap engineering—to the realization of robust quantum error correction.

Science & Technology - Other Topics↗

Characterization of MKIDs for CMB Observation at 220 GHz With the South Pole Telescope

We present an updated design of the 220 GHz microwave kinetic inductance detector (MKID) pixel for SPT-3G+, the next-generation camera for the South Pole Telescope. We show results of the dark testing of a 63-pixel array with mean inductor quality factor $Q_{i} = 4.8 \times 10^{5}$, aluminum inductor transition temperature $T_{c} = 1.19$K, and kinetic inductance fraction $\alpha _{k} = 0.32$. We optically characterize both the microstrip-coupled and CPW-coupled resonators, and find both have a spectral response close to prediction with an optical efficiency of $\eta \sim 70\%$. However, we find slightly lower optical response on the lower edge of the band than predicted, with neighboring dark detectors showing more response in this region, though at level consistent with less than 5% frequency shift relative to the optical detectors. The detectors show polarized response consistent with expectations, with a cross-polar response of $\sim 10\%$for both detector orientations.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

X-ray Computed Tomography Data of Dense Metallic Components

The data shared in here are X-ray computed tomography (XCT) scans of a hexagonal fuel nozzle in 3 sections with the Metrotom 800 system at the Manufacturing Demonstration Facility (MDF) at Oak Ridge National Laboratory. The data are used in the paper "Tomographic Sparse View Selection using the View Covariance Loss, by Lin et al. (doi:10.1109/TPAMI.2025.36000720), accepted to the international conference on computational imaging (ICCP 2025). Figures 4-7 in the paper describe the part/XCT scan. File name Descriptions: Bottom section: TCR- Single Channeled SRC L 2019-3-18 12-26-41.hdf5 Medium section: TCR- Single Channeled SRC M 2019-3-18 13-8-9.hdf5 Top section: TCR- Single Channeled SRC T 2019-3-18 13-45-39.hdf5 Each hdf5 file contains projection data, and all the relevant X-ray CT scan setting. The full list of included attributes: distance_unit: Units of all distances specified angle_unit : Units of the angles angles: Array of all angles used voxel_size_xy: Baseline recon (if any) has this voxel size in the in-plane direction voxel_size_z: Baseline recon (if any) has this voxel size in the cross-plane direction det_pixel_size_col: Size of the detector pixels in the column dimension det_pixel_size_row: Size of the detector pixels in the row dimension src_iso_dist: Source to iso-center distance iso_det_dist: Iso-center to detector distance det_angle: If the detector is rotated/tilted, this angle corresponds to that value det_row_offset: Center of rotation offset in the vertical direction det_col_offset: Center of rotation offset in the horizontal direction reconstruction: A baseline reconstruction stored as 3D array BHC params: Beam-hardening parameters - Van De Casteel Model - if it has been used to pre-process the projections We also provided a python script (hdf_io.py) that allows the user to read the relevant data from each hdf5 file.

Ziabari, Amir [Oak Ridge National Laboratory]↗

Characterization of a SiPM-based monolithic neutron scatter camera using dark counts

The Single Volume Scatter Camera (SVSC) Collaboration aims to develop portable neutron imaging systems for a variety of applications in nuclear non-proliferation. Conventional double-scatter neutron imagers are composed of several separate detector volumes organized in at least two planes. A neutron must scatter in two of these detector volumes for its initial trajectory to be reconstructed. As such, these systems typically have a large footprint and poor geometric efficiency. We report on the design and characterization of a prototype monolithic neutron scatter camera that is intended to significantly improve upon the geometrical shortcomings of conventional neutron cameras. The detector consists of a 50 mm×56 mm× 60 mm monolithic block of EJ-204 plastic scintillator instrumented on two faces with arrays of 64 Hamamatsu S13360-6075PE silicon photomultipliers (SiPMs). The electronic crosstalk is limited to < 5% between adjacent channels and < 0.1% between all other channel pairs. SiPMs introduce a significantly elevated dark count rate over PMTs, as well as correlated noise from after-pulsing and optical crosstalk. In this article, we characterize the dark count rate and optical crosstalk and present a modified event reconstruction likelihood function that accounts for them. We find that the average dark count rate per SiPM is 4.3 MHz with a standard deviation of 1.5 MHz among devices. The analysis method we employ to measure internal optical crosstalk also naturally yields the mean and width of the single-electron pulse height. Here, we calculate separate contributions to the width of the single-electron pulse-height from electronic noise and avalanche fluctuations. We demonstrate a timing resolution for a single-photon pulse to be (128 ± 4) ps. Finally, coincidence analysis is employed to measure external (pixel-to-pixel) optical crosstalk. We present a map of the average external crosstalk probability between 2×4 groups of SiPMs, as well as the in-situ timing characteristics extracted from the coincidence analysis. Further work is needed to characterize the performance of the camera at reconstructing single- and double-site interactions, as well as image reconstruction.

47 OTHER INSTRUMENTATION↗

DarkQuest: A dark sector upgrade to SpinQuest at the 120 GeV Fermilab Main Injector

Expanding the mass range and techniques by which we search for dark matter is an important part of the worldwide particle physics program. Accelerator-based searches for dark matter and dark sector particles are a uniquely compelling part of this program as a way to both create and detect dark matter in the laboratory and explore the dark sector by searching for mediators and excited dark matter particles. This paper focuses on developing the DarkQuest experimental concept and gives an outlook on related enhancements collectively referred to as LongQuest. DarkQuest is a proton fixed-target experiment with leading sensitivity to an array of visible dark sector signatures in the MeV-GeV mass range. Because it builds off of existing accelerator and detector infrastructure, it offers a powerful but modest-cost experimental initiative that can be realized on a short timescale.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

MARTA’s (Muon Array with RPC for Tagging Air showers) DAQ system

The successful development of a low-power and standalone DAQ system to be used with RPCs is presented. These will be used to measure the muonic component of cosmic-ray’s air showers. The system includes a front-end, high voltage, detector monitoring and a Central Unit to manage the different components of the system. The front-end is based on the MAROC ASIC coupled to an FPGA respecting the strict demands of field operations in cosmic-ray experiments. Prototypes were produced and deployed, performing as expected. An engineering array will be installed at the Pierre Auger Observatory.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Demonstration and performance of an online data selection algorithm for liquid argon time projection chambers using MicroBooNE

The MicroBooNE detector is a liquid argon time projection chamber (LArTPC) that produces three-dimensional images of particle interactions using ionization charge collected by anode wire plane arrays and scintillation light collected by a light detection system. In addition to testing long-standing experimental neutrino anomalies and performing measurements of neutrino interactions with argon nuclei using the Fermilab Booster Neutrino Beam, MicroBooNE aims to develop methodologies for rare beyond the Standard Model and off-beam physics searches. Looking ahead to the upcoming Deep Underground Neutrino Experiment (DUNE), with MicroBooNE serving as a valuable testbed, achieving high sensitivity and livetime for off-beam physics while satisfying data processing and storage constraints will require data-driven, intelligent, and online or real-time data selection techniques. These techniques are essential for reducing data rates and preserving rare signals with high accuracy. In this paper, we describe a fast data selection algorithm suitable for online execution to identify electrons from stopping cosmic ray muons in the MicroBooNE detector utilizing ionization charge information, and present its performance. This represents the first demonstration of online data selection in a LArTPC using real data and charge information exclusively and provides an important proof-of-principle for applying such techniques to other LArTPC experiments such as the Short-Baseline Near Detector and DUNE.

Abratenko, P. [Tufts U. (main)]↗

Evidence of Coherent Elastic Neutrino-Nucleus Scattering with COHERENT’s Germanium Array

We report the first detection of coherent elastic neutrino-nucleus scattering (CEvNS) on natural germanium, measured at the Spallation Neutron Source at Oak Ridge National Laboratory. The Ge-Mini detector of the COHERENT collaboration employs large-mass, low-noise, high-purity germanium spectrometers, enabling excellent energy resolution, and an analysis threshold of 1.5 keV electron-equivalent ionization energy. We observe an on-beam excess of 20.6$^{+7.1}_{−6.3}$ counts with a total exposure of 10.22 GWhkg, and we reject the no-CEvNS hypothesis with 3.9⁢𝜎 significance. The result agrees with the predicted standard model of particle physics signal rate within 2⁢𝜎.

Electroweak interaction↗

Characterization of the DUNE photodetectors and study of the event burst phenomenon

The Deep Underground Neutrino Experiment (DUNE) is an upcoming neutrino physics experiment that will answer some of the most compelling questions in particle physics and cosmology. The DUNE far detectors employ silicon photomultipliers (SiPMs) to detect light produced by charged particles interacting in a large liquid argon time projection chamber (LArTPC). The SiPMs are photosensors consisting of an array of single-photon avalanche diodes (SPAD) operating in Geiger mode. Their high sensitivity and dynamic range, as well as the possibility to fill large surfaces with high-granularity sensors, makes them an ideal choice for DUNE. An international consortium of research groups is currently engaged in a systematic comparison of the performances of the SiPM models that have been custom developed for DUNE by two manufacturers. Such detailed studies, which include gain measurements and a structure study of the dark count rate at 77 K, are meant to determine the best choice of the photodetection system for DUNE, as well as characterize the response of the chosen detectors for the DUNE simulation. Moreover, an investigation of a newly observed phenomenon, in which quick bursts of tens of events at close range are collected in individual SiPMs, is being carried out, which potentially impacts the design of future models and their implementation in particle physics experiments.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

An energy-dependent electro-thermal response model of CUORE cryogenic calorimeter

The Cryogenic Underground Observatory for Rare Events (CUORE) is the most sensitive experiment searching for neutrinoless double-beta decay (0νββ) in 130 Te. CUORE uses a cryogenic array of 988 TeO 2 calorimeters operated at ~10 mK with a total mass of 741 kg. To further increase the sensitivity, the detector response must be well understood. Here, we present a non-linear thermal model for the CUORE experiment on a detector-by-detector basis. We have examined both equilibrium and dynamic electro-thermal models of detectors by numerically fitting non-linear differential equations to the detector data of a subset of CUORE channels which are well characterized and representative of all channels. Further, we demonstrate that the hot-electron effect and electric-field dependence of resistance in NTD-Ge thermistors alone are inadequate to describe our detectors' energy-dependent pulse shapes. We introduce an empirical second-order correction factor in the exponential temperature dependence of the thermistor, which produces excellent agreement with energy-dependent pulse shape data up to 6 MeV. We also present a noise analysis using the fitted thermal parameters and show that the intrinsic thermal noise is negligible compared to the observed noise for our detectors.

47 OTHER INSTRUMENTATION↗

Online and Offline Analytical Techniques to Quantify Chloride Salts

This report serves as the deliverable for Milestone- M3FT-26AN080502017: Summary of Accomplishments for Online and Offline Analytical Qualification of Chloride Fuel Salts. The qualification of chloride fuel salts is a critical challenge for the development and deployment of molten salt reactors (MSRs), requiring precise control and verification of chemical and isotopic composition to ensure predictable neutronics, manageable radiological behavior, and safe plant operation. Currently, there is no universally accepted standard for fuel salt qualification, particularly for chloride-based systems, which underscores the need for robust, science-based analytical methodologies. This report presents a comprehensive evaluation of both online and offline techniques for quantifying chloride fuel salts, including multielectrode array voltammetry, differential thermal analysis (DTA) probes, inert gas fusion (IGF) analyzers, and gamma spectrometry using the Mirion NAIS-2x2 NaI(Tl) detector. The integration of these methods enables real-time monitoring of actinide concentrations, redox state, and thermophysical properties, as well as high-precision measurement of impurities and isotopic composition. The acquisition of advanced instruments such as the Bruker Leonardo G6 for IGF and the Mirion NaI detector enhances analytical capabilities, supporting the establishment of operational envelopes and impurity thresholds. These approaches provide essential data for neutronic impact evaluation, feedstock documentation, and compliance with nuclear safety standards. The methodologies developed and validated in this report lay the groundwork for future standardized protocols, bridging the gap between laboratory research and commercial reactor operation, and advancing the safe and efficient deployment of chloride-fueled MSRs.

Polke, Amber↗

Development of selenium imaging detectors toward a definitive search for neutrinoless ββ decay

Imaging sensors made from an ionization target layer of amorphous selenium (aSe) coupled to a silicon complementary metal-oxide-semiconductor (CMOS) active pixel array for charge readout are a promising technology for neutrino physics. The high spatial resolution in a solid-state target provides unparalleled rejection of backgrounds from natural radioactivity in the search for the neutrinoless ββ decay of 82 Se. With this grant, we experimentally determined two crucial factors necessary for the realization of this novel technology: i) the intrinsic energy response of aSe to ionizing electrons, and ii) the feasibility to couple a target layer of aSe to a CMOS active pixel sensor to image electron tracks.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

DR-RINS: Digital real-time relative intensity noise suppressor for pump–probe spectroscopy and microscopy

Relative intensity noise (RIN) inherent in fiber lasers poses a serious obstacle to their use in pump–probe spectroscopy and imaging. RIN can be removed through an analog balanced detector, or, as we have previously shown, software adaptive noise cancellation (ANC) on digitized signals. One major drawback to software ANC is the added time required for digitizing and post-processing. In this article, we describe a design for ANC on a field-programmable gate array (FPGA), making use of high-level synthesis tools and fixed-point arithmetic to achieve real-time laser RIN suppression at 25 MHz sample rates. Unlike the software-ANC approach, the FPGA-ANC device can serve as a dedicated drop-in denoiser, placed between the detectors and a commercial lock-in amplifier. We demonstrate its application to transient absorption spectroscopy and microscopy, lowering the noise floor to ~17 dB above the shot noise limit. Furthermore, we demonstrate a dramatic improvement in data acquisition time from ~6 h to ~5 min in a real-time imaging scenario.

47 OTHER INSTRUMENTATION↗

Reaction losses of charged particles in CsI(Tl) crystals

To efficiently detect energetic light charged particles, it is common to use arrays of energy-loss telescopes involving two or more layers of detection media. As the energy of the particles increases, thicker layers are usually needed. However, carrying out measurements with thick-telescopes may require corrections for the losses due to nuclear reactions induced by the incident particles on nuclei within the detector and for the scattering of incident particles out of the detector, without depositing their full energy in the active material. In this paper, we develop a method for measuring such corrections and determine the reaction and out-scattering losses for data measured with the silicon-CsI(Tl) telescopes of the newly developed HiRA10 array. Here, the extracted efficiencies are in good agreement with model predictions using the GEANT4 reaction loss algorithm for Z = 1 and Z = 2 isotopes. After correcting for the HiRA10 geometry, we obtain a general function that describes the loss of efficiency due to reaction losses in CsI(Tl) crystals as a function of range.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗