Calculation and Verification of a Planar Pencil Beam Kernel Through the Hankel Transform of Measured OARs for a Radiosurgery System with Cones
A planar multienergetic pencil beam kernel with rotational symmetry is calculated for a stereotactic radiosurgery system, SRS, BrainLAB with cones, employing the deconvolution method of the off axis ratio profile, OAR, corresponding to the cone of 35 mm in diameter for a 6 MV photon beam produced by a linear accelerator Varian 2100 C/D. Before the deconvolution, the experimental OAR is corrected for beam divergence and variations of the spectral fluence {Phi}, using a boundary function BF. The function BF and the fluence {Phi} are transformed to the conjugate space with the zero order Hankel function, which is the appropriate transform due to the radial symmetry of the circular beams generated by the cones. The kernel in the conjugate space is obtained as the ratio of the transform of BF to the transform of {Phi}, therefore the kernel in the real space is calculated as the inverse transform of the kernel in the conjugate space. To validate the kernel in the real space, it is convolved with the fluence of the cones of 7.5, 12.5, 15, 17.5, 20, 22.5, 25, 30 and 35 mm in diameter. The comparison of the OARs calculated and measured shows a maximum difference of 4.5% in the zones of high gradient of dose, and a difference less than 2% in the regions of low gradient of dose. Finally, the expanded uncertainty of the kernel is estimated and reported.