We undertook, at Brookhaven National Laboratory (BNL), full 3D simulations of one-sided dual-type column 3D Si detectors (p-type substrate) after irradiation levels up to 1 x 10{sup 16} n{sub eq}/cm{sup 2}. Etching does not extend all the way through the substrate, so that all electrodes are on the front side; the backside was neither supported nor processed, covered only with a layer of silicon dioxide. Hence, BNL's one-sided dual-type column 3D detectors are true one-sided detectors. Our simulations of the electric fields at various fluences will add to our knowledge about the practicable applicability of these detectors in the high-radiation environment of future colliders. Simulations show that the full depletion voltage, V{sub fd}, for a dual-column 3D detector is about 1.4 times higher than that of a 2D (planar) pad detector with a thickness d, the same as the column spacing, L{sub p}, in the 3D detector. Moreover, the highest E-field is near the n{sup +} column, while the high E-field mainly is distributed between the n{sup +} and p{sup +} columns. The low E-field is between the two p{sup +} columns, with its lowest point in the center of the unit cell with two p{sup +} columns and two n{sup +} columns. To fully deplete a dual-column 3D detector at 1 x 10{sup 16} n{sub eq}/cm{sup 2} with a reasonable bias ({le} 200 V) necessitates reducing the column spacing, L{sub p}, to 30 {mu}m. The volume under the columns (10% of the total volume) can be depleted with a modest bias ({le}200 V); this underlying volume is not a dead volume. In addition, this depletion volume can support detection sensitivity directly under the columns, thereby lowering the effective dead volume in a 3D Si detector.