Excerpt from Oxygen Monitor Clean Testing Report Performed in 2010
The Rosemount X-STREAM Series X2GP is a paramagnetic oxygen analyzer with a 0-25% oxygen measurement range. It claims a drift less than 2% of span (0.10% volume oxygen) per week, and a response time of less than 5 seconds (for the analyzer to report 90% of a step change). Our test equipment was set-up such that the same gas was running through each oxygen analyzer at the same time. This allowed Calibration Error, Calibration Drift, and Loss of Power Testing to be run on all monitors at the same time, eliminating any variability between the analyzers from tolerances in the gas cylinder, ambient temperature, ambient pressure, test duration, etc. An APTech pressure regulator was placed on the gas cylinder manifold and a 23 psia Paroscientific pressure transducer measured the regulator’s outlet pressure. This did not give the inlet pressure on any one analyzer but outlet pressure from the manifold before the stream was split. The APTech pressure regulator was used to keep operating conditions consistent. A DWYER rotameter was installed on each oxygen monitor’s outlet piping. With the exception of Variation of Flow Rate Testing, the rotameter was left fully open for all testing. The flow rate through each oxygen monitor was controlled by a Swagelok needle valve that was installed on each analyzer’s inlet piping. Additionally, a filter was installed in the each monitor’s inlet piping to keep particles from the measurement cells and a Swagelok ball valve was installed on each monitor’s outlet piping to isolate each monitor between tests. No pumps were installed. Each of the cylinders used was at a pressure greater than ambient room barometric pressure. This pressure differential was used to move gas through the testing equipment. The flow rate through each analyzer was controlled by the Swagelok needle valve installed on each analyzer’s inlet piping. Finally, for all tests, a Paroscientific Digiquartz Barometer was used to measure ambient room barometric pressure and a Fluke Thermo-hygrometer was used to measure ambient room temperature.