Comparison of two Calibrations of NASA's MC60 Five-component Semi-span Balance
Two separate manual calibrations of NASA’s MC60 five–component semi–span balance were compared. The calibrations were performed in 1999 and 2019using CALSPAN Force Measurement System’s Large Load Rig. The Iterative Method was chosen as the load prediction algorithm and data reduction matrices were generated from both calibration data sets. Then, the applied calibration load schedules, the standard deviations of the calibration load residuals, the prime sensitivities of the five gages, and the maximum outputs at capacity of the five gages were compared. In addition, the calibration data of 1999 was used as check load data for the data reduction matrix that was obtained from the 2019 data. Overall, the agreement between the data analysis results is very good considering the facts that two different load schedules were used for the calibrations, different people performed the calibrations, different instrumentation was used to measure the loads and the outputs, the Large Load Rig was moved to a different site in between the calibrations, and no tare load corrections could be applied to the 1999 data set. Nevertheless, the calibration data of 2019 is clearly the better data set. This observation is no surprise considering that CALSPAN Force Measurement Systems made significant investments in calibration hardware and process improvements during the last 20 years.