The Suzaku High Resolution X-ray Spectrometer

The high resolution X-Ray Spectrometer (XRS) has been designed to provide the Suzaku Observatory with very high spectral resolution, non-dispersive spectroscopy from 0.3 to 12 keV. This energy range encompasses the most diagnostically-rich part of the x-ray band. The sensor consists of a 32 channel array of x-ray of microcalorimeters, each with an energy resolution of about 6 eV. The very low temperature required for operation of the array (60 mK) is provided by a four-stage cooling system containing a single stage ADR, superfluid He Cryostat, solid Ne Dewar, and a single-stage Stirling-cycle cooler. The Suzaku/XRS is the first orbiting x-ray microcalorimeter spectrometer and has been designed to last more than three years in orbit. The early verification phase of the mission demonstrated that the instrument was working properly and that the cryogen consumption rate was low enough to ensure a mission lifetime exceeding 3 years. However, the liquid He cryogen was completely vaporized two weeks after opening the dewar guard vacuum vent. The problem has been traced to inadequate venting of the dewar He and Ne gases out of the spacecraft into space. In this paper we present the design of the XRS instrument and describe the in-flight performance.

Kelley, Richard L.↗

Update on the Wide-field Infrared Survey Explorer (WISE)

The Wide-field Infrared Survey Explorer (WISE), a NASA MIDEX mission, will survey the entire sky in four bands from 3.3 to 23 microns with a sensitivity 1000 times greater than the IRAS survey. The WISE survey will extend the Two Micron All Sky Survey into the thermal infrared and will provide an important catalog for the James Webb Space Telescope. Using 1024(sup 2) HgCdTe and Si:As arrays at 3.3, 4.7, 12 and 23 microns, WISE will find the most luminous galaxies in the universe, the closest stars to the Sun, and it will detect most of the main belt asteroids larger than 3 km. The single WISE instrument consists of a 40 cm diamond-turned aluminum afocal telescope, a two-stage solid hydrogen cryostat, a scan mirror mechanism, and reimaging optics giving 5 resolution (full-width-half-maximum). The use of dichroics and beamsplitters allows four color images of a 47' x47' field of view to be taken every 8.8 seconds, synchronized with the orbital motion to provide total sky coverage with overlap between revolutions. WISE will be placed into a Sun-synchronous polar orbit on a Delta 7320-10 launch vehicle. The WISE survey approach is simple and efficient. The three-axis-stabilized spacecraft rotates at a constant rate while the scan mirror freezes the telescope line of sight during each exposure. WISE has completed its mission Preliminary Design Review and its NASA Confirmation Review, and the project is awaiting confirmation from NASA to proceed to the Critical Design phase. Much of the payload hardware is now complete, and assembly of the payload will occur over the next year. WISE is scheduled to launch in late 2009; the project web site can be found at www.wise.ssl.berkeley.edu.

aster↗

The Development and Mission of the Space Infrared Telescope Facility

This paper provides an overview of the SIRTF mission, telescope, cryostat, instruments, spacecraft, orbit, operations and project management approach; and this paper serves as an introduction to the accompanying set of detailed papers about specific aspects of SIRTF.

Spitzer Space Telescope↗

Gradiometer Using Middle Loops as Sensing Elements in a Low-Field SQUID MRI System

A new gradiometer scheme uses middle loops as sensing elements in lowfield superconducting quantum interference device (SQUID) magnetic resonance imaging (MRI). This design of a second order gradiometer increases its sensitivity and makes it more uniform, compared to the conventional side loop sensing scheme with a comparable matching SQUID. The space between the two middle loops becomes the imaging volume with the enclosing cryostat built accordingly.

Penanen, Konstantin↗

NASA Tech Briefs, February 2010

Topics covered include: Insulation-Testing Cryostat With Lifting Mechanism; Optical Testing of Retroreflectors for Cryogenic Applications; Measuring Cyclic Error in Laser Heterodyne Interferometers; Self-Referencing Hartmann Test for Large-Aperture Telescopes; Measuring a Fiber-Optic Delay Line Using a Mode-Locked Laser; Reconfigurable Hardware for Compressing Hyperspectral Image Data; Spatio-Temporal Equalizer for a Receiving-Antenna Feed Array; High-Speed Ring Bus; Nanoionics-Based Switches for Radio-Frequency Applications; Lunar Dust-Tolerant Electrical Connector; Compact, Reliable EEPROM Controller; Quad-Chip Double-Balanced Frequency Tripler; Ka-Band Waveguide Two-Way Hybrid Combiner for MMIC Amplifiers; Radiation-Hardened Solid-State Drive; Use of Nanofibers to Strengthen Hydrogels of Silica, Other Oxides, and Aerogels; Two Concepts for Deployable Trusses; Concentric Nested Toroidal Inflatable Structures; Investigating Dynamics of Eccentricity in Turbomachines; Improved Low-Temperature Performance of Li-Ion Cells Using New Electrolytes; Integrity Monitoring of Mercury Discharge Lamps; White-Light Phase-Conjugate Mirrors as Distortion Correctors; Biasable, Balanced, Fundamental Submillimeter Monolithic Membrane Mixer; ICER-3D Hyperspectral Image Compression Software; and Context Modeler for Wavelet Compression of Spectral Hyperspectral Images.

Source record↗

Infrared Spectra, Index of Refraction, and Optical Constants of Nitrile Ices Relevant to Titan's Atmosphere

Spectra and optical constants of nitrite ices known or suspected to be in Titan's atmosphere are presented from 2.5 to 200 microns (4000 to 50 per cm ). These results are relevant to the ongoing modeling of Cassini CIRS observations of Titan's winter pole. Ices studied include: HCN, hydrogen cyanide; C2N2, cyanogen; CH3CN, acetonitrile; C 2H5CN, propionitrile; and HC3N, cyanoacetylene. For each of these molecules we report new measurements of the index of refraction, n, determined in both the amorphous- and crystallinephase at 670 nm. Spectra were measured and optical constants were calculated for each nitrite at a variety of temperatures including 20, 35, 50, 75, 95, and 110 K, in the amorphous- and crystalline-phase. This laboratory effort uses a dedicated FTIR spectrometer to record transmission spectra of thin-film ice samples. Laser interference is used to measure film thickness during condensation onto a transparent cold window attached to the tail section of a closed-cycle helium cryostat. Optical constants, real (n) and imaginary (k) refractive indices, are determined using Kramers-Kronig (K-K) analysis. Our calculation reproduces the complete spectrum, including all interference effects. Index of refraction measurements are made in a separate dedicated FTIR spectrometer where interference deposit fringes are measured using two 670 nm lasers at different angles to the ice substrate. A survey of these new measurements will be presented along with a discussion of their validation, errors, and application to Titan data.

Moore, Marla↗

Apparatus for Measuring Total Emissivity of Small, Low-Emissivity Samples

An apparatus was developed for measuring total emissivity of small, lightweight, low-emissivity samples at low temperatures. The entire apparatus fits inside a small laboratory cryostat. Sample installation and removal are relatively quick, allowing for faster testing. The small chamber surrounding the sample is lined with black-painted aluminum honeycomb, which simplifies data analysis. This results in the sample viewing a very high-emissivity surface on all sides, an effect which would normally require a much larger chamber volume. The sample and chamber temperatures are individually controlled using off-the-shelf PID (proportional integral derivative) controllers, allowing flexibility in the test conditions. The chamber can be controlled at a higher temperature than the sample, allowing a direct absorptivity measurement. The lightweight sample is suspended by its heater and thermometer leads from an isothermal bar external to the chamber. The wires run out of the chamber through small holes in its corners, and the wires do not contact the chamber itself. During a steady-state measurement, the thermometer and bar are individually controlled at the same temperature, so there is zero heat flow through the wires. Thus, all of sample-temperature-control heater power is radiated to the chamber. Double-aluminized Kapton (DAK) emissivity was studied down to 10 K, which was about 25 K colder than any previously reported measurements. This verified a minimum in the emissivity at about 35 K and a rise as the temperature dropped to lower values.

Tuttle, James↗

Design of Multilayer Insulation for the Multipurpose Hydrogen Test Bed

Multilayer insulation (MLI) is a critical component for future, long term space missions. These missions will require the storage of cryogenic fuels for extended periods of time with little to no boil-off and MLI is vital due to its exceptional radiation shielding properties. Several MLI test articles were designed and fabricated which explored methods of assembling and connecting blankets, yielding results for evaluation. Insight gained, along with previous design experience, will be used in the design of the replacement blanket for the Multipurpose Hydrogen Test Bed (MHTB), which is slated for upcoming tests. Future design considerations are discussed which include mechanical testing to determine robustness of such a system, as well as cryostat testing of samples to give insight to the loss of thermal performance of sewn panels in comparison to the highly efficient, albeit laborious application of the original MHTB blanket.

Marlow, Weston A.↗

Thermal Analysis of Low Layer Density Multilayer Insulation Test Results

Investigation of the thermal performance of low layer density multilayer insulations is important for designing long-duration space exploration missions involving the storage of cryogenic propellants. Theoretical calculations show an analytical optimal layer density, as widely reported in the literature. However, the appropriate test data by which to evaluate these calculations have been only recently obtained. As part of a recent research project, NASA procured several multilayer insulation test coupons for calorimeter testing. These coupons were configured to allow for the layer density to be varied from 0.5 to 2.6 layer/mm. The coupon testing was completed using the cylindrical Cryostat-l00 apparatus by the Cryogenics Test Laboratory at Kennedy Space Center. The results show the properties of the insulation as a function of layer density for multiple points. Overlaying these new results with data from the literature reveals a minimum layer density; however, the value is higher than predicted. Additionally, the data show that the transition region between high vacuum and no vacuum is dependent on the spacing of the reflective layers. Historically this spacing has not been taken into account as thermal performance was calculated as a function of pressure and temperature only; however the recent testing shows that the data is dependent on the Knudsen number which takes into account pressure, temperature, and layer spacing. These results aid in the understanding of the performance parameters of MLI and help to complete the body of literature on the topic.

Johnson, Wesley L.↗

Thermal Performance of Low Layer Density Multilayer Insu1ation Using Liquid Nitrogen

In order to support long duration cryogenic propellant storage, the Cryogenic Fluid Management (CFM) Project of the Exploration Technology Development Program (ETDP) is investigating the long duration storage propertie$ of liquid methane on the lunar surface. The Methane Lunar Surface Thermal Control (MLSTC) testing is using a tank of the approximate dimensions of the Altair ascent tanks inside of a vacuum chamber to simulate the environment in low earth orbit and on the lunar surface. The thermal performance testing of multilayer insulation (MLI) coupons that are fabricated identically to the tank applied insulation is necessary to understand the performance of the blankets and to be able to predict the performance of the insulation prior to testing. This coupon testing was completed in Cryostat-100 at the Cryogenics Test Laboratory. The results showed the properties of the insulation as a function of layer density, number of layers, and warm boundary temperature. These results aid in the understanding of the performance parameters o fMLI and help to complete the body of literature on the topic.

Johnson, Wesley L.↗

The Detector Subsystem for the SXS Instrument on the Astro-H Observatory

The Soft X-ray Spectrometer (SXS) instrument on the Astro-H observatory is based on a 36 pixel x-ray calorimeter array cooled to 50 mK in a sophisticated spaceflight cryostat. The SXS is a true spatial-spectral instrument, where each spatially discrete pixel functions as a high-resolution spectrometer. Here we discuss the SXS detector subsystem that includes the detector array, the anticoincidence detector, the first stage amplifiers, the thermal and mechanical staging of the detector, and the cryogenic bias electronics. The design of the SXS detector subsystem has significant heritage from the Suzaku/XRS instrument but has some important modifications that increase performance margins and simplify the focal plane assembly. Notable improvements include x-ray absorbers with significantly lower heat capacity, improved load resistors, improved thermometry, and a decreased sensitivity to thermal radiation. These modifications have yielded an energy resolution of 3.5-4.0 eV FWHM at 6 keV for representative devices in the laboratory, giving considerable margin against the 7 eV instrument requirement. We expect similar performance in flight

Porter, Frederick↗

NASA Tech Briefs, May 2004

Topics include: Embedded Heaters for Joining or Separating Plastic Parts; Curing Composite Materials Using Lower-Energy Electron Beams; Aluminum-Alloy-Matrix/Alumina-Reinforcement Composites; Fibrous-Ceramic/Aerogel Composite Insulating Tiles; Urethane/Silicone Adhesives for Bonding Flexing Metal Parts; Scalable Architecture for Multihop Wireless ad Hoc Networks; Improved Thermoplastic/Iron-Particle Transformer Cores; Cooperative Lander-Surface/Aerial Microflyer Missions for Mars Exploration Dual-Frequency Airborne Scanning Rain Radar Antenna System Eight-Channel Continuous Timer Reduction of Phase Ambiguity in an Offset-QPSK Receiver Ambient-Light-Canceling Camera Using Subtraction of Frames Lightweight, Flexible, Thin, Integrated Solar-Power Packs Windows(Registered Trademark)-Based Software Models Cyclic Oxidation Behavior Software for Analyzing Sequences of Flow-Related Images Improved Ball-and-Socket Docking Mechanism Two-Stage Solenoid Ordered Nanostructures Made Using Chaperonin Polypeptides Low-Temperature Plasma Functionalization of Carbon Nanotubes Improved Cryostat for Cooling a Wide Panel Current Pulses Momentarily Enhance Thermoelectric Cooling Hand-Held Color Meters Based on Interference Filters Calculating Mass Diffusion in High-Pressure Binary Fluids Fresnel Lenses for Wide-Aperture Optical Receivers Increasing Accuracy in Computed Inviscid Boundary Conditions Higher-Order Finite Elements for Computing Thermal Radiation Radar for Monitoring Hurricanes from Geostationary Orbit Time-Transfer System for Two Orbiting Spacecraft

Source record↗

Optical Position Encoders for High or Low Temperatures

Optoelectronic pattern-recognition systems (optical encoders) for measuring positions of objects of interest at temperatures well below or well above room temperature are undergoing development. At present, the development effort is concentrated on absolute linear, rotary, and Cartesian encoders and Cartesian autocollimators for scientific instruments that operate in cryostats.

Leviton, Douglas B.↗

NASA Tech Briefs, March 2003

Topics covered include: Tool for Bending a Metal Tube Precisely in a Confined Space; Multiple-Use Mechanisms for Attachment to Seat Tracks; Force-Measuring Clamps; Cellular Pressure-Actuated Joint; Block QCA Fault-Tolerant Logic Gates; Hybrid VLSI/QCA Architecture for Computing FFTs; Arrays of Carbon Nanotubes as RF Filters in Waveguides; Carbon Nanotubes as Resonators for RF Spectrum Analyzers; Software for Viewing Landsat Mosaic Images; Updated Integrated Mission Program; Software for Sharing and Management of Information; Update on Integrated Optical Design Analyzer; Optical-Quality Thin Polymer Membranes; Rollable Thin Shell Composite-Material Paraboloidal Mirrors; Folded Resonant Horns for Power Ultrasonic Applications; Touchdown Ball-Bearing System for Magnetic Bearings; Flux-Based Deadbeat Control of Induction-Motor Torque; Block Copolymers as Templates for Arrays of Carbon Nanotubes; Throttling Cryogen Boiloff To Control Cryostat Temperature; Collaborative Software Development Approach Used to Deliver the New Shuttle Telemetry Ground Station; Turbulence in Supercritical O2/H2 and C7H16/N2 Mixing Layers; and Time-Resolved Measurements in Optoelectronic Microbioanal.

Source record↗

NASA Tech Briefs, April 2003

Topics include: Tool for Bending a Metal Tube Precisely in a Confined Space; Multiple-Use Mechanisms for Attachment to Seat Tracks; Force-Measuring Clamps; Cellular Pressure-Actuated Joint; Block QCA Fault-Tolerant Logic Gates; Hybrid VLSI/QCA Architecture for Computing FFTs; Arrays of Carbon Nanotubes as RF Filters in Waveguides; Carbon Nanotubes as Resonators for RF Spectrum Analyzers; Software for Viewing Landsat Mosaic Images; Updated Integrated Mission Program; Software for Sharing and Management of Information; Optical-Quality Thin Polymer Membranes; Rollable Thin Shell Composite-Material Paraboloidal Mirrors; Folded Resonant Horns for Power Ultrasonic Applications; Touchdown Ball-Bearing System for Magnetic Bearings; Flux-Based Deadbeat Control of Induction-Motor Torque; Block Copolymers as Templates for Arrays of Carbon Nanotubes; Throttling Cryogen Boiloff To Control Cryostat Temperature; Collaborative Software Development Approach Used to Deliver the New Shuttle Telemetry Ground Station; Turbulence in Supercritical O2/H2 and C7H16/N2 Mixing Layers; and Time-Resolved Measurements in Optoelectronic Microbioanal.

Source record↗

The High Resolution Microcalorimeter Soft X-Ray Spectrometer for the Astro-H Mission

We are developing the Soft X-Ray Spectrometer for the JAXA Astro-H mission. The instrument features a 5 eV, 36-pixel array of micro calorimeters designed for high spectral resolution from 0.3-12 keV at the focus of an x-ray mirror, providing a field of view of3 x 3 arcmin. The principal components of the spectrometer are the microcalorimeter detector system, a 3-stage ADR and dewar. The dewar is a long-life, hybrid design with a superfluid He cryostat, Joule-Thomson cooler, and Stirling coolers. We describe the present design of the SXS instrument and initial engineering model test results.

Kelley, Richard L.↗

Stray Light Suppression in the Goddard IRAM 2-Millimeter Observer (GISMO)

The Goddard-IRAM Superconducting 2 Millimeter Observer (GISMO) is an 8xl6 Transition Edge Sensor (TES) array of bolometers built as a pathfinder for TES detector development efforts at NASA Goddard Space Flight Center. GISMO has been used annually at the Institut de Radioastronomie Millimetrique (IRAM) 30 meter telescope since 2007 under engineering time and was opened in the spring of 2012 to the general astronomical community. The spring deployment provided an opportunity to modify elements of the room temperature optics before moving the instrument to its new permanent position in the telescope receiver cabin. This allowed for the possibility to extend the cryostat, introduce improved cold baffling and thus further optimize the stray light performance for final astronomical use of the instrument, which has been completed and validated. We will demonstrate and discuss several of the methods used to quantify and limit the influence of stray light in the GISMO camera.

Sharp, E. H.↗

Experimental Results of Integrated Refrigeration and Storage System Testing

Launch operations engineers at the Kennedy Space Center have identified an Integrated Refrigeration and Storage system as a promising technology to reduce launch costs and enable advanced cryogenic operations. This system uses a close cycle Brayton refrigerator to remove energy from the stored cryogenic propellant. This allows for the potential of a zero loss storage and transfer system, as well and control of the state of the propellant through densification or re-liquefaction. However, the behavior of the fluid in this type of system is different than typical cryogenic behavior, and there will be a learning curve associated with its use. A 400 liter research cryostat has been designed, fabricated and delivered to KSC to test the thermo fluid behavior of liquid oxygen as energy is removed from the cryogen by a simulated DC cycle cryocooler. Results of the initial testing phase focusing on heat exchanger characterization and zero loss storage operations using liquid oxygen are presented in this paper. Future plans for testing of oxygen densification tests and oxygen liquefaction tests will also be discussed. KEYWORDS: Liquid Oxygen, Refrigeration, Storage

Notardonato, W. U.↗