Evaluation of high resolution X-ray monochromatic wavelength dispersive X-ray fluorescence instrument for determining Pu and U in nuclear reprocessing streams
In this study, an alpha-prototype benchtop instrument was developed and tested for measuring Pu and U in spent fuel dissolver tanks in a nuclear fuel reprocessing plant. The high resolution X-ray (hiRX) instrument was designed with the aim of providing a new capability to measure samples with high radiation background with selectivity, accuracy, and precision using ultra-low volumes (7 μL) of liquid. The instrument was tested with specimens from the operational reprocessing facility at Savannah River National Laboratory. The results showed that bias was <6% for Pu nitrate process solution and < 18% for Pu in U/HNO 3 matrix. Instrumental precision was <1% RSD while between microcell repeatability was <6% RSD. The limit of detection for Pu in HNO 3 matrix was 3 × 10 -5 g/L (100 s, 35 W). Finally, this study demonstrates that hiRX technology has the potential to provide rapid and simultaneous quantification of Pu and U without sample preparation, making it suitable for process monitoring and materials control and accountability applications.