Measuring tropospheric HNO3 - Problems and prospects for Nylon filter and mist chamber techniques
A series of laboratory and field measurements was performed to evaluate the mist chamber technique for determining tropospheric HNO3 concentrations. Both the mist chamber and standard Nylon filter techniques exhibit high collection efficiency and excellent agreement measuring HNO3 vapors from a permeation source. When simultaneously sampling ambient air in eastern Virginia, the Nylon filter measured an average of 70 percent higher HNO3 concentration than the mist chamber technique. The results indicate that O3 causes a low-level positive artifact interference in HNO3 measurements performed with the filter technique. This O3-induced error is small, however, compared to the large difference between atmospheric HNO3 concentrations determined with the two techniques. It is hypothesized that unidentified (organic?) nitrogen species in the atmosphere react for form NO3(-) on the filter and this phenomenon may interfere with Nylon filter measurements of HNO3 vapor. These potential interferences did not appear to affect measurements of HNO3 with the mist chamber method.