TU-H-206-03: Characterizing B1 Inhomogeneities in DCE MRI
Purpose: Dynamic Contrast Enhanced (DCE) MRI is a valuable technique for measuring perfusion and permeability characteristics of tumors. Exogenous contrast concentrations are calculated based on changes in T{sub 1} measured using fast 3D gradient echo (FLASH) sequences. However, the slab selective pulses used in 3D MRI may result in B{sub 1} inhomogeneities across the volume of interest that can lead to errors in T{sub 1} and thus the estimated gadolinium concentration. We compared three FLASH DCE sequences (GRE, TWIST, and VIBE) to determine their signal homogeneity across slices and the accuracy in calculating T{sub 1} using acquisitions with variable flip angles. Methods: The sequences were tested at 3 T on a Siemens mMR (VB20P) using a doped water phantom 3.75 g/L NiSO{sub 4} - 6H{sub 2}O + 5 g/L NaCl (T{sub 1} = 104 ms) and a 2% agar, 0.67% NaCl phantom (T{sub 1}= 1.71 s). 2D EPI B{sub 1} maps and inversion recovery T{sub 1}maps were acquired for ground truth. 3D MRI was acquired at different flip angles to generate a T{sub 1} map. Regions of interest were drawn to measure signal inside the phantoms as a function of slice position. The T{sub 1} for each slice ROI was fit to the FLASH steady-state model of magnetization. Results: Based on the data, GRE gave the most uniform signal homogeneity and T{sub 1} values in the middle slices of the 3D volume. The 3D VIBE sequence had the largest region of signal inhomogeneity compared to the 3D GRE and TWIST sequences. VIBE’s B{sub 1} inhomogeneity is inconsistent at low flip angles. However, VIBE resulted in more slices with T{sub 1} values similar to the ground truth. Conclusion: The central 1/3 of the slices yielded signals that result in T{sub 1} fits consistent with the ground truth. However, the remaining slices required some form of B{sub 1} inhomogeneity correction for quantitative DCE analysis. The research was supported in part by NIH NCI Grant R01CA159471.