A comprehensive final summary of the MCNP calculations performed concerning the UW 137 Cs release event
This document combines and supersedes earlier reports (e.g. LA-UR-19-32358 and LA-UR-20- 22216) that discussed MCNP-calculated conversion factors (count rate per unit activity per unit area) for collimated 2”x2” NaI detectors (Ludlum model 44-10 and Thermo model SPA3) in contact with various surfaces of interest to the University of Washington 137 Cs release event. These surfaces include: Exterior surface of round and rectangular ducts, Exterior surface of outdoor brick work, Exterior surface of cast iron pipes, Exterior surface of an O 2 filter assembly For each surface, the calculated conversion factors allowed inferences on 137 Cs activity on interior surfaces based on exterior contact measurements of the 662 keV photopeak count rate. The MCNP calculations included the effect of attached collimators (i.e. either a Pb/Sb-based Ludlum or a W-based “Rees” model) on detector response. As a consequence, the conversion factors only apply to the appropriate collimated 2”x2” NaI detector operated in pulse height analysis (PHA) mode. Other assumptions affecting the calculations and valid application of the conversion factors will be outlined below.