A New Small Nano-Kelvin Resolution Thermometer for Low Temperature Experiments
A small, high resolution paramagnetic susceptibility thermometer was developed using a GdCl3 paramagnetic salt. The device uses a SQUID (Superconducting Quantum Interference Device) magnetometer to determine the temperature dependent magnetization of the salt in a magnetic field. The magnetic field is provided by a pair of small samarium cobalt permanent magnet disks situated inside the thermometer housing. The eliminates the need for a heavy, charging solenoid used in a conventional SQUID based magnetic thermometer system. This thermometer can resolve approximately 10 (exp -9)K near the liquid-gas critical point of He-3(approximately 3.31K). The drift rate of the thermometer was measured to be less than 2 x 10 (exp -13) K per second. This light weight thermometer (approximately 7 g) is a candidate for use in future low temperature Space Shuttle and International Space Station experiments. The present thermometer design easily permits a variety of magnetic materials for use in other applications.