Ex-situ Heat Treatment of TEM Foils in a Custom Titanium Fixture: A Case Study on Ni-based Superalloy
The advancement of microstructural characterization at the nanoscale is critical to understanding the microstructural evolution and performance of engineering materials. Transmission electron microscopy (TEM) plays a vital role in such investigations, particularly when coupled with controlled ex- and in-situ experiments. In this study, we introduce a novel method for ex-situ heat treatment (HT) of TEM foils using a custom-designed fixture, made from pure titanium, and vacuum furnace to inhibit oxidation. This approach ensures precise temperature control and minimal sample contamination during HT, critical for substructural characterization and analysis of metallic materials. The fixture consists of a titanium base with three slits, for three sample foils, and a titanium cap for closure. The entire assembly is placed in a vacuum heat treatment furnace with high vacuum capability to prevent oxidation. To validate the effectiveness of the setup, precipitation behavior and microstructural changes were studied in an IN725 variant heat-treated at 500°C for 1 hour and 282 variant heat-treated at 700°C for 1 hour, as case studies.