Results from the Balloon Ozone Intercomparison Campaign (BOIC)
In the Balloon Ozone Intercomparison Campaign (BOIC), several in situ UV absorption photometers, two solar UV absorption photometers, electrochemical sondes, and a mass spectrometer were intercompared in three flight missions. Concurrent data from Umkehr and satellite observations are also intercompared. The National Bureau of Standards provided a 'standard' ozone source for intercomparing the in situ instruments and ground pressure. Preliminary findings indicate that the standard deviation of the sensitivities among 17 instruments against the NBS reference was about 11 percent. These differences appear in flight at the lower levels and change at higher altitudes, indicating height-dependent errors. The difference among five in-situ UV photometers flown together ranged by plus or minus 8 percent during ascent to about 41 km. During float at 42 km, the difference nearly doubled. During descent, the difference decreased to about 4 percent, which is much closer to the expected accuracy of these instruments. Results from UV solar radiometers have been systematically higher than those from UV photometers by 15 to 20 percent - a very important disagreement that needs to be resolved.