Detector Technologies for Sub-500μm High-Sensitivity PET Imaging via a Novel PET Insert Approach
The objective of this project was to develop detector technologies that would enable an ultrahigh resolution Virtual Pinhole (VP) PET insert device to provide sub-500 μm resolution high-sensitivity PET imaging of a mouse in the future. To achieve this goal, we proposed to develop and characterize finely pixelated cadmium zinc telluride (CZT) detectors and the associated readout electronics with the following specific aims: 1. Develop pixelated CZT detectors with 350μm pitches using 2-5 mm substrates; characterize their spatial, energy and timing performance through experiments; and optimize the anode design with steering grid if found necessary. 2. Develop a high-bandwidth readout system using a novel ASIC that can be directly bonded to CZT detectors with 2048 anodes of 350μm pitches; optimize its overall performance for VP-PET applications considering the tradeoffs between spatial resolution (in 3D), count rate capability, timing and energy resolutions. 3. Evaluate the performance of a VP-PET insert based on the proposed detector technology through Monte Carlo simulation and experimental validation. Overall, we have completed all three specific aims and demonstrated that pixelated CZT detectors of 350μm pitches, combined with VP-PET geometry, can provide PET image resolution of ~460 μm FWHM for small animal imaging applications.