A new active base of photomultiplier R4125 designed for the PbWO4 calorimeter

Abstract

This paper presents the design, performance, and results of the radiation tests of an active base for Hamamatsu R4125 photomultiplier tube. The active base was designed at Jefferson Lab and comprises of a high voltage divider and an on-board amplifier. The photomultiplier with the active base is used to detect light from lead tungstate scintillating crystals of the forward electromagnetic calorimeter of the GlueX detector. The active base amplifier allows to operate the tube at lower high voltage and thus to limit the photomultiplier anode current to a few micro ampereres at the maximum counter rate of 1 MHz counter, while retaining the dynamic range of output signals. The performance of calorimeter modules instrumented with the active base was studied using detector prototypes positioned into the beam of photons. The key performance parameters such as the linearity, high-rate capability, and the energy resolution verified that the active base design meets the detector specifications

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Somov, Alexander [Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)], Barbosa, Fernando [Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)], Berdnikov, Vladimir [Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)] (ORCID:0000000316034320), Popov, Vladimir [Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)]. 2025-10-30. A new active base of photomultiplier R4125 designed for the PbWO4 calorimeter. https://doi.org/10.2172/3000424

Cite the original work for its findings. Save a collection to share your selection of sources.