An MLC calibration method using a detector array

Abstract

Purpose: The authors have developed a quantitative calibration method for a multileaf collimator (MLC) which measures individual leaf positions relative to the MLC backup jaw on an Elekta Synergy linear accelerator. Methods: The method utilizes a commercially available two-axis detector array (Profiler 2; Sun Nuclear Corporation, Melbourne, FL). To calibrate the MLC bank, its backup jaw is positioned at the central axis and the opposing jaw is retracted to create a half-beam configuration. The position of the backup jaws field edge is then measured with the array to obtain what is termed the radiation defined reference line. The positions of the individual leaf ends relative to this reference line are then inferred by the detector response in the leaf end penumbra. Iteratively adjusting and remeasuring the leaf end positions to within specifications completes the calibration. Using the backup jaw as a reference for the leaf end positions is based on three assumptions: (1) The leading edge of an MLC leaf bank is parallel to its backup jaw's leading edge, (2) the backup jaw position is reproducible, and (3) the measured radiation field edge created by each leaf end is representative of that leaf's position. Data from an electronic portal imaging device (EPID) were used in a similar analysis to check the results obtained with the array. Results: The relative leaf end positions measured with the array differed from those measured with the EPID by an average of 0.11 {+-}0.09 mm per leaf. The maximum leaf positional change measured with the Profiler 2 over a 3 month period was 0.51 mm. A leaf positional accuracy of {+-}0.4 mm is easily attainable through the iterative calibration process. The method requires an average of 40 min to measure both leaf banks. Conclusions: This work demonstrates that the Profiler 2 is an effective tool for efficient and quantitative MLC quality assurance and calibration.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Simon, Thomas A., Kahler, Darren, Simon, William E., Fox, Christopher, Li, Jonathan, Palta, Jatinder, Liu, Chihray. 2009-10-15. An MLC calibration method using a detector array. https://doi.org/10.1118/1.3218767

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Performance evaluation of neutron noise analysis in detecting special nuclear materials

Reliable and rapid inspection techniques play a vital role in preventing illicit trafficking of special nuclear materials. Active interrogation systems using neutrons produced by portable, high-flux deuterium-deuterium or deuterium-tritium neutron generators are being actively developed as a secondary scanning tool for this purpose. In this study, a neutron noise analysis-based approach for detecting unshielded and shielded special nuclear materials by using a pulsed deuterium-tritium neutron generator was evaluated. Here, this approach analyzes the fluctuation of neutron counts. Its performance was quantified with regard to time-to-detection to achieve a minimum probability of detection of 99% and a probability of false alarm of less than 1% considering various amounts of special nuclear materials and different shielding configurations. It was demonstrated that this approach could detect 17 uranium slugs in 5 s given a neutron generator yield of 8.1 × 10 7 n/s. These slugs could be detected within a reasonable time frame (200 s) when they were shielded by 10.16 cm of high-density polyethylene. The results obtained using the neutron noise analysis approach were compared with those obtained using the commonly used differential die-away analysis technique, a sensitive technique for detecting the presence of fissile materials by utilizing the prompt fission neutrons produced when the source neutrons from a neutron generator are completely diminished. For example, the time to detect 2 unshielded uranium slugs was 2.1 s when using the differential die-away analysis technique; it increased to 93 s for the neutron noise analysis approach. Although the noise analysis-based approach exhibits an overall performance which is not as good as that of differential die-away, neutron noise provides an alternative method for effective detection of special nuclear materials.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

A proximity-focusing RICH detector for the ePIC Experiment at the EIC

The Electron-Proton/Ion Collider Experiment (ePIC) will be a large, multi-purpose detector to be installed at the Electron-Ion Collider (EIC) being built at Brookhaven National Laboratory. As robust particle identification (PID) capabilities are essential for fully realizing the EIC science program, ePIC contains several PID subsystems spanning different angular ranges. PID capability in the electron-going endcap is provided by a proximity-focusing Ring Imaging Cherenkov detector (pfRICH) designed to deliver at least 3σ separation between pions and kaons for momenta up to 7 GeV/ c . It will also aid with electron-hadron discrimination at low momentum and assist in the determination of the collision time (t 0 ). This contribution will summarize the design of the pfRICH as well as ongoing fabrication and component testing efforts. geant-based performance simulations will also be discussed.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Design of a high voltage delivery system for noble liquid time projection chambers

Noble liquid time projection chambers (TPCs) are a leading technology in the detection of ionizing radiation, particularly in applications such as accelerator neutrino physics, dark matter detection, and neutrinoless double beta decay. This work addresses the design considerations for implementing stable high voltage (HV) systems within large noble liquid TPCs, with a focus on the nEXO experiment. Utilizing insights from prior HV research and experimental investigations, we outline factors influencing HV stability and discuss design choices to improve stability and prevent electrical discharges. A novel HV delivery system concept is presented, tailored for the nEXO TPC, which incorporates these design considerations while also meeting the stringent radiopurity requirements of the nEXO neutrinoless double beta decay search. These design considerations and their specific implementation towards a HV delivery system offer guidance to future experiments applying high voltage in noble liquid environments.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗