Abstract We have implemented an “Acute Shortage Ventilator” (ASV) motivated by the COVID-19 pandemic and the possibility of severe ventilator shortages in the near future. The unit cost per ventilator is less than $400 US excluding the patient circuit parts. The ASV mechanically compresses a self-inflating bag resuscitator, uses a modified patient circuit, and is commanded by a microcontroller and an optional laptop. It operates in both Volume-Controlled Assist-Control mode and a Pressure-Controlled Assist-Control mode. It has been tested using an artificial lung against the EURS guidelines. The key design goals were to develop a simple device with high performance for short-term use, made primarily from common hospital parts and generally-available non-medical components, and at low cost and ease in manufacturing.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Akerib, D. S., Ames, A. (ORCID:000000028249714X), Breidenbach, M. (ORCID:0000000150220835), Bressack, M., Breur, P. A. (ORCID:0000000153975299), Charles, E. (ORCID:0000000239257802), Gaba, D., Herbst, R., Ignarra, C. M. (ORCID:0000000283771366), Luitz, S., Miller, E. H., Mong, B., Shutt, T. A., Wittgen, M.. 2020-11-19. Acute Shortage Ventilator. https://doi.org/10.1101/2020.07.20.20158147