Modular total absorption spectrometer
Abstract
The design and performance of the Modular Total Absorption Spectrometer built and commissioned at the Oak Ridge National Laboratory is presented here. The active volume of the detector is approximately one ton of NaI(Tl), which results in very high full γ energy peak efficiency of 71% at 6 MeV and nearly flat efficiency of around 81.5% for low energy γ-rays between 300 keV and 1 MeV. In addition to the high peak efficiency, the modular construction of the detector permits the use of a γ-coincidence technique in data analysis as well as β-delayed neutron observation.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Karny, Marek, Rykaczewski, Krzysztof P. (ORCID:0000000247726330), Fijałkowska, Aleksandra G., Rasco, Bertis C., Wolińska-Cichocka, Marzena, Grzywacz, R. K., Goetz, K. C., Miller, D., Zganjar, E. F.. 2016-08-24. Modular total absorption spectrometer. https://doi.org/10.1016/j.nima.2016.08.046