DOE OSTI · 1483460
Acoustic Chamber Characterization
Abstract
The basis of this project was to characterize the various uncertainty contributors of an acoustic chamber. The acoustic chamber will be used to calibrate and characterize infrasound sensors used in the field in the frequency range of 0.001 Hz to 4 Hz. The components characterized include the internal volume of the chamber, the piston area of the speaker creating the dynamic sound wave, the environmental stabilization of the chamber and the chamber's leak rate. Also, the resonant frequency of the chamber was evaluated and found to be far outside the frequency band of interest.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
O'Brien, Edward, Kypta, Timothy, Stanford, Joshua, Tran, Hy D.. 2018-11-01. Acoustic Chamber Characterization. https://doi.org/10.2172/1483460. https://engineeringpapers.com/record/osti%3A1483460/acoustic-chamber-characterization