NASA NTRS · 20150008534
Polymer-Based Carbon Monoxide Sensors
Abstract
Polymer-based sensors have been used primarily to detect volatile organics and inorganics; they are not usually used for smaller, gas phase molecules. We report the development and use of two types of polymer-based sensors for the detection of carbon monoxide. Further understanding of the experimental results is also obtained by performing molecular modeling studies to investigate the polymer-carbon monoxide interactions. The first type is a carbon-black-polymer composite that is comprised of a non-conducting polymer base that has been impregnated with carbon black to make it conducting. These chemiresistor sensors show good response to carbon monoxide but do not have a long lifetime. The second type of sensor has a non-conducting polymer base but includes both a porphyrin-functionalized polypyrrole and carbon black. These sensors show good, repeatable and reversible response to carbon monoxide at room temperature.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Homer, M. L., Shevade, A. V., Zhou, H., Kisor, A. K., Lara, L. M., Yen, S.-P. S., Ryan, M. A.. 2010-11-01. Polymer-Based Carbon Monoxide Sensors. https://ntrs.nasa.gov/citations/20150008534
Cite the original work for its findings. Save a collection to share your selection of sources.